OSS-B1

User Manual

Summary of Contents for OSS-B1

Page 1: ...OSS B1 User Manual ...

Page 2: ... 1 Unpacking your OSS B1 Current Meter 7 Part III Operation 8 Part IV Maintenance 10 4 1 Service and Recalibration 11 Part V Troubleshooting 12 Part VI Repair 14 6 1 Replacement of Cable Ends 14 6 2 Part List 17 Part VII Technical Data 25 Part VIII Obligations of the Operator and Disposal 26 8 1 Obligations of the Operator 26 8 2 Dismantling Disposal 26 ...

Page 3: ...ormation provided in this manual was deemed accurate as of the publication date However updates to this information may have occurred This manual does not include all of the details of design production or variation of the equipment nor does it cover every possible situation which may arise during installation operation or maintenance HyQuest Solutions shall not be liable for any incidental indire...

Page 4: ...Fan A additional fans available see ch Technical Data 3 m connecting lead tools oil spare bearings spare reed switch assembly Rod 1 8 m 3 0 6 m 20 mm diameter in canvas carry bag with point and base plate Carry case durable waterproof Pelican Case 2022 HyQuest Solutions 4 25 ...

Page 5: ...found before starting to work as well as for defects found while working Do not use the HyQuest Solutions OSS B1 in areas where there is a danger of explosion The present user manual specifies environmental climatic operating conditions as well as mechanical and electrical conditions Installation wiring powering up and operating the HyQuest Solutions OSS B1 must strictly comply with these specific...

Page 6: ...winches and permanent cableway installations with streamlined or ground feeling weights A tail fin assembly is provided to control directional stability during suspended measurements The OSS B1 comes with lots of accessories which enable the user to endure water flow measurement easily and accurately All current meters manufactured by Hyquest Solutions are calibrated in our purpose built high velo...

Page 7: ...SS B1 Body OSS B1 Lead Body Key Fan type A Fan type 1 2 and 4 are accessories 3 wading rods 1 8m 3 sections or 3 00m 3 sections CMC20A PVD100 or CMC3 counter 2 spare bearings 1 reed switch Screwdriver Short Screwdriver Small O Rings for fan carrier 1 reed switch cover 2 lead 1 terminal post 1 Post ground 1 oil bottle Rating Tables Note Tail Fin Stem Assembly and other accessories are supplied on r...

Page 8: ...w resistance signal passage The voltage must not exceed 9V DC Provided the impulse counters have spark suppression circuits the meter can be loaded with up to 1 6 watts All our counters meet these requirements refer to specification Four type of stainless steel Fans 4 can be used with this meter all these Fans are supported on bearings which run within the oil filled hub of the Fan assembly The ca...

Page 9: ... 2022 HyQuest Solutions 9 Operation Figure 2 Figure 3 Figure 4 ...

Page 10: ...m 110 sec 1 Minimum 85 sec 2 Minimum 70 sec 4 Minimum 90 sec After the spin test is performed connect your current meter to the counter using the lead this important to test the lead reed switch and counter Section 2 During and after Discharge Measurements After every discharge measurement or one hour of use the current meter should be cleaned and refilled with oil if continuing a discharge measur...

Page 11: ... the ends connections If the current meter fails the spin test refer to the mechanical fault finding section for details Test lead counter and reed switch see electrical fault finding section for details For more information see the following subsection Service and Recalibration 4 1 Service and Recalibration Recalibrate the current meter every 300 hours of use or once a year whichever comes first ...

Page 12: ...to ON and INT ie no time limit Spin the Fan and if the audible signal sounds continuously the fault is a CLOSED CIRCUIT If no audible signal is heard the fault then is an OPEN CIRCUIT Closed Circuit Fault Symptom Possible Cause Action Fault occurs only when meter in water Water entering contact plug receptacle and earth between plug and body Replace damaged plug or missing seal see figure 7 Circui...

Page 13: ...ure 3 Check fit of plug in hole Remove wipe surface clean and check spring compression Re assemble with silicone grease on all surfaces No closure of reed switch contact when magnet in position Damaged Reed Switch Assy Replace Reed switch operating but signal not passing to body Faulty earth contact 12 Figure 3 Check spring compression Remove assembly and wipe clean contact surfaces with silicon g...

Page 14: ...ons within the return paperwork This will assist our test engineers with their diagnosis 3 Please do not ship the goods prior to obtaining the reference number HyQuest Solutions will not reject any equipment that arrives without reference number however it may take us longer to process Custom requirements for items sent to HyQuest Solutions for warranty or non warranty repairs Check with your nati...

Page 15: ... heat shrink sleeve 3 red and black identifying sleeves 4 and plug covers 5 d Push soldered conductors into plugs 6 and lock with grubscrew 7 e Bring cover forward over plugs f Position heat shrink sleeve centrally over grey cable insulation and fix in place Note Repair to either plug end individually can be made also by only cutting away the damaged end and following steps b d e Figure 5 ...

Page 16: ... back into plug holder and screw plug cap into place f Fit rubber seal 12 to plug pin g For the negative end slide on black heat shrink tube 13 Push conductor into the earth pin 14 and solder in place Bring heat shrink sleeve forward onto earth pin and fix down with heat h Fix down heat shrink sleeve 4 over main cable termination i Repairs can be made individually to the positive end with steps 2 ...

Page 17: ... 2022 HyQuest Solutions 17 Repair 6 2 Part List ...

Page 18: ... 2022 HyQuest Solutions 18 Repair ...

Page 19: ... 2022 HyQuest Solutions 19 Repair ...

Page 20: ... 2022 HyQuest Solutions 20 Repair ...

Page 21: ... 2022 HyQuest Solutions 21 Repair ...

Page 22: ...te WRS04 1 Canvas Carrying Bag WR11 1 Fixing Screw SAN20 04 5 Screwdriver SC002 06 1 Three W R series wading rod sections plated alloy steel graduations each 2cm and markings at decimetre 10cm intervals 0 1 metre 1 2 metre 2 3 metre WR01 WR02 WR03 1 1 1 Point WR04 01 1 Foot Plate WR04 02 1 Canvas Carrying Bag WR11L 1 Fixing Screw SAN20 04 5 Screwdriver SC002 06 1 METER RELOCATING DEVICE STAINLESS ...

Page 23: ... 2022 HyQuest Solutions 23 Repair Universal Wading Rods Relocating Devices ...

Page 24: ... 2022 HyQuest Solutions 24 Repair Gauging Rod Stand ...

Page 25: ...r Specifications The choice of fan propeller depends on the specifics of the site such as discharge and the geometry of the riverbed or channel Please contact us for advice Fan No Diam Pitch Start Velocity m sec Max Velocity m sec Component Effect A 100 mm 0 125 m 0 025 5 45 1 125 mm 0 250 m 0 025 10 5 2 125 mm 0 500 m 0 040 10 5 4 80 mm 0 125 m 0 025 4 5 Packed Dimensions Mass basic kit Kit in ca...

Page 26: ...ections must be observed 8 2 Dismantling Disposal When disposing of the units and their accessories the applicable local regulations regarding environment disposal and occupational safety must be observed Before dismantling Electrical Devices Switch off the units Disconnect electrical appliances from the power supply regardless of whether the appliances are connected to the mains or to another pow...

Page 27: ...are committed to supporting efforts to avoid and recycle waste Please consider Avoidance before recycling Recycling before disposal This symbol indicates that the scrapping of the unit must be carried out in accordance with Directive 2012 19 EU Please observe the local implementation of the directive and any accompanying or supplementary laws and regulations ...

Page 28: ...RS Group HyQuest Solutions KISTERS LATAM HyQuest Solutions NZ Ltd HyQuest Solutions Pty Ltd HyQuest Solutions KISTERS AG 1 561 459 4876 sales hsa kisters net 57 350 575 4079 sales latam hyquestsolutions com 64 7 857 0810 sales hyquestsolutions co nz 612 9601 2022 sales hyquestsolutions com au 49 2408 9385 0 info hyquestsolutions eu ...

Page 29: ...www hyquestsolutions com 2022 HyQuest Solutions 21 04 2022 ...