Keysight NFA Series Noise Figure Analyzers Service Guide
313
Power Supply/Midplane Troubleshooting
A7 Midplane Description
Power Supply Dithering
A triangle wave of approximately 100 Hz is generated and goes directly to the
A6 Power Supply assembly. This is used to frequency modulate the power
supply switching frequency for the purpose of lowering any power supply
related interference. If for some reason the level of the power supply related
interference is higher than normal this circuitry, or the A6 Power Supply
assembly, could be suspect.
Miscellaneous Circuitry
Other circuitry on the A7 Midplane include a circuit to phase lock the CPU
clock to the analyzer's 10 MHz reference and a Complex Programmable Logic
Device (CPLD) used as an Instrument Local Bus (ILB) buffer to A16 Reference
Assembly connector at J21.
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...