Keysight NFA Series Noise Figure Analyzers Service Guide
83
Instrument Messages
Event Messages
–310
System Error; No
supported source
found
Signal source at given IP address is not responding / IP does not belong to a
source. Check IP address and network connection
–310
System Error; source
connection lost,
check interface
connection
Signal source at given IP address is not responding / IP does not belong to a
source. Check IP address and network connection
–310
System error; The
configured PTP
hardware driver could
not be instantiated.
The PTP driver failed on initialization
–310
System error; The PTP
hardware driver
reported a
configuration error.
Failure in the execution of the PTP driver. The most likely cause of this error is a
mismatch between versions of the PTP driver and the LXI middleware
–310
System error; The PTP
ordinary clock
reported a
configuration error.
Failure in execution of the DMC libraries’ PTP controller
–310
System error; The
Trigger alarm delayed
LAN event was not
scheduled due to an
existing pending
event.
Delayed LAN events cannot occur too close together (within 20 ms)
–310
System error; The
Trigger alarm delayed
LAN event was not
scheduled due to a
conflict with an
existing scheduled
alarm.
Delayed LAN events cannot occur too close to a scheduled Alarm (within 20
ms)
–310
System error; The
Trigger alarm was not
scheduled due to a
conflict with an
existing scheduled
alarm.
Alarms cannot be scheduled to happen too close together (within 20 ms)
–311
Memory error
There is a physical problem with the instrument memory, such as a parity error
–312
PUD memory lost
Protected user data saved by the *PUD command has been lost
–313
Calibration memory
lost
The nonvolatile calibration data used by the *CAL? command has been lost
Err#
Message
Verbose/Correction Information
Содержание N8973B
Страница 4: ...4 ...
Страница 44: ...44 Keysight NFA Series Noise Figure Analyzers Service Guide Overview How to Return Your Instrument for Service ...
Страница 132: ...132 Keysight NFA Series Noise Figure Analyzers Service Guide Instrument Messages Condition Messages ...
Страница 228: ...228 Keysight NFA Series Noise Figure Analyzers Service Guide RF Section Troubleshooting N8976B Troubleshooting ...
Страница 268: ...268 Keysight NFA Series Noise Figure Analyzers Service Guide Analog Digital IF Troubleshooting 25 MHz BW IF Section ...
Страница 322: ...322 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Страница 332: ...332 Keysight NFA Series Noise Figure Analyzers Service Guide Block Diagrams Block Diagrams NFA Computer Block Diagram ...
Страница 350: ...350 Keysight NFA Series Noise Figure Analyzers Service Guide Service and Diagnostics Menus Front Panel Key Test ...
Страница 470: ...470 Keysight NFA Series Noise Figure Analyzers Service Guide Assembly Replacement Procedures Front Frame Assembly ...