Performance Verification
7
Average Gain of Zoomed-In Channel
N2820/1A User’s Guide
99
positive gain: ________________
7
Set the DC power supply to –20 mV.
8
Connect the multimeter on the DC power supply as shown in
9
Record the value of –V
in
measured on the multimeter:
–V
in
:________________
10
Connect the multimeter’s leads to the banana plug on the output
of the N1022A/B as shown in
11
Record the value of –V
out
measured on the multimeter:
–V
out
:________________
12
Calculate the negative gain of the zoomed-in (primary) channel
and record the result in the following line:
negative gain:________________
13
Calculate the average of the two gains:
positive gain
+V
out
+V
in
------------
=
NOTE
Although this is the recommended setting, you can use any negative
voltage less than –35 mV.
negative gain
–V
out
–V
in
------------
=
Содержание N2820A
Страница 1: ...Keysight N2820 1A High Sensitivity Current Probes User s Guide...
Страница 30: ...30 N2820 1A User s Guide 2 Making Measurements Quick Start Figure 10 R Sense Entry Box...
Страница 74: ...74 N2820 1A User s Guide 4 Probe Calibration To Calibrate the Probe on InfiniiVision Oscilloscopes...
Страница 84: ...84 N2820 1A User s Guide 5 Characteristics and Specifications Dimensions...
Страница 94: ...94 N2820 1A User s Guide 6 Performance Data Plots Input Load Model...
Страница 106: ...106 N2820 1A User s Guide 7 Performance Verification Performance Test Record...
Страница 109: ...N2820 1A User s Guide 109...
Страница 110: ...This information is subject to change without notice Keysight Technologies 2018 March 2018 N2820 97006 www keysight com...