70
N2820/1A User’s Guide
4
Probe Calibration
To Calibrate the Probe on InfiniiVision Oscilloscopes
To Calibrate the Probe on InfiniiVision Oscilloscopes
The following test equipment are needed for the calibration
procedure.
Table 9
Required Calibration Test Equipment
Calibration Procedure
1. Connect the probe to the oscilloscope
a
Turn the oscilloscope on.
b
Plug the primary channel of the N2820A probe into the
oscilloscope channel you are intending to use.
c
If applicable, plug the secondary channel of the probe into the
oscilloscope channel you are intending to use.
d
Configure the zoomed-in path to the primary or secondary
channel as necessary.
e
Solder the sensor leads onto the N2825A user-defined head as
f
Attach the N2825A user-defined head to the probe.
Test Equipment
Description
N2825A User-Defined
Probe Head
Provided with probe.
Pomona 3781-12-2 Red
Minigrabber Test Clip
Patch Cord
http://www.pomonaelectronics.com/
for
more information and how to order this
product.
Pomona 3781-12-0 Black
Minigrabber Test Clip
Patch Cord
http://www.pomonaelectronics.com/
for
more information and how to order this
product.
Содержание N2820A
Страница 1: ...Keysight N2820 1A High Sensitivity Current Probes User s Guide...
Страница 30: ...30 N2820 1A User s Guide 2 Making Measurements Quick Start Figure 10 R Sense Entry Box...
Страница 74: ...74 N2820 1A User s Guide 4 Probe Calibration To Calibrate the Probe on InfiniiVision Oscilloscopes...
Страница 84: ...84 N2820 1A User s Guide 5 Characteristics and Specifications Dimensions...
Страница 94: ...94 N2820 1A User s Guide 6 Performance Data Plots Input Load Model...
Страница 106: ...106 N2820 1A User s Guide 7 Performance Verification Performance Test Record...
Страница 109: ...N2820 1A User s Guide 109...
Страница 110: ...This information is subject to change without notice Keysight Technologies 2018 March 2018 N2820 97006 www keysight com...