72
N2820/1A User’s Guide
4
Probe Calibration
To Calibrate the Probe on InfiniiVision Oscilloscopes
ii) Turn on Averaging with
16 averages
.
iii) For both channels, set the V/div as small as possible to fit the
waveform on screen. For the zoomed-in channel, this should be
less than
10
μ
A
. For the zoomed-out channel, this should be less
than
2mA
. If the waveform is not on screen at these settings,
increase the V/div and adjust the offset until these settings can
be used.
iv) Turn on
Vavg
measurements for both channels.
d
To set the zoomed-out offset, turn
R2
using the supplied
screwdriver until the Vavg measurement is
+/- 1mA
.
e
To set the zoomed-in offset, turn
R4
using the supplied
screwdriver until the Vavg measurement is
+/- 1mA
.
Figure 34
Location of Calibration Adjustments
f
Press the Channel on/off key to turn the channel on (if the
channel is off).
NOTE
The R2 and R4 adjustments are located on the underside of the probe
. The screwdriver provided with the
probe is the preferred adjustment tool.
Содержание N2820A
Страница 1: ...Keysight N2820 1A High Sensitivity Current Probes User s Guide...
Страница 30: ...30 N2820 1A User s Guide 2 Making Measurements Quick Start Figure 10 R Sense Entry Box...
Страница 74: ...74 N2820 1A User s Guide 4 Probe Calibration To Calibrate the Probe on InfiniiVision Oscilloscopes...
Страница 84: ...84 N2820 1A User s Guide 5 Characteristics and Specifications Dimensions...
Страница 94: ...94 N2820 1A User s Guide 6 Performance Data Plots Input Load Model...
Страница 106: ...106 N2820 1A User s Guide 7 Performance Verification Performance Test Record...
Страница 109: ...N2820 1A User s Guide 109...
Страница 110: ...This information is subject to change without notice Keysight Technologies 2018 March 2018 N2820 97006 www keysight com...