102
N2820/1A User’s Guide
7
Performance Verification
Average Gain of Zoomed-Out Channel
3
Record the value of +V
in
measured on the multimeter:
+V
in
:________________
4
Move the multimeter’s leads to the banana plug on the output of
.
5
Record the value of +V
out
measured on the multimeter:
+V
out
:________________
Figure 52
Test Setup for Measuring V
out
6
Calculate the positive gain of the zoomed-out (secondary) channel
and record the result in the following line: