Keysight N2820/1A High-Sensitivity Current Probes
User’s Guide
5 Characteristics and
Specifications
Current Range Graphs / 78
Zoomed-In Input Voltage Range Versus Supply Voltage / 80
Zoomed-In Input Current Range Versus Supply Voltage / 81
Dimensions / 83
This chapter provides the characteristics and specifications for the N2820A and
N2821A probes. The probe should be warmed up for at least 20 minutes before
any testing and the environmental conditions should not exceed the probe’s
specified limits.
Table 10 Characteristics and Specifications
Item
Characteristic
Bandwidth (–3 dB)
Zoom-Out Channel: DC to 3 MHz
Zoom-In Channel: DC to 500 kHz
Risetime (Tr = 0.35/BW, 10% – 90%)
Zoom-Out Channel: < 0.116 µs
Zoom-In Channel: < 0.7 µs
Minimum Measurable Current
250 µA (with N2822A 20 m
Ω
,)
50 µA (with N2824A 100 m
Ω
,)
5 mA (with N2825A user defined 1 m
Ω
,)
100 nA (with N2825A user defined 1 k
Ω
,)
a
Maximum Measurable Current
5A (with N2822A 20 m
Ω
, 500 mW)
2.2A (with N2824A 100 m
Ω
, 500 mW)
5A (with N2825A, 1 m
Ω
installed)
1.2 mA (with N2825A user defined 1 k
Ω
,)
Содержание N2820A
Страница 1: ...Keysight N2820 1A High Sensitivity Current Probes User s Guide...
Страница 30: ...30 N2820 1A User s Guide 2 Making Measurements Quick Start Figure 10 R Sense Entry Box...
Страница 74: ...74 N2820 1A User s Guide 4 Probe Calibration To Calibrate the Probe on InfiniiVision Oscilloscopes...
Страница 84: ...84 N2820 1A User s Guide 5 Characteristics and Specifications Dimensions...
Страница 94: ...94 N2820 1A User s Guide 6 Performance Data Plots Input Load Model...
Страница 106: ...106 N2820 1A User s Guide 7 Performance Verification Performance Test Record...
Страница 109: ...N2820 1A User s Guide 109...
Страница 110: ...This information is subject to change without notice Keysight Technologies 2018 March 2018 N2820 97006 www keysight com...