Using N2825A Probe Heads
3
To Add a Custom Internal R
SENSE
Resistor
N2820/1A User’s Guide
63
Figure 29
Location of R
SENSE
Resistor (R1)
5
Replace the head’s cover using the four screws. (M1.6 x 0.35, 5 mm
long, Keysight part number 0515-5210)
6
Connect the N2825A to an N2820/1A probe and connect the probe
to the oscilloscope.
7
On Infiniium oscilloscopes,
a
Click the
Setup
>
Probe Configuration
menu command.
b
Select the tab for the probe’s primary channel. N2821A probes
only have a primary channel.
c
In the
User Def Head Resistance
field, enter the resistance of the
DUT’s
R
SENSE
resistor in ohms.
d
In the dialog box, click
Add Head
to give your head with
associated external
R
SENSE
resistor a unique name.
CAUTION
If you are not skilled in soldering, ask a trained assembler to change the
resistor.
CAUTION
To avoid damaging the aluminum casting, do not over tighten the screws.
Содержание N2820A
Страница 1: ...Keysight N2820 1A High Sensitivity Current Probes User s Guide...
Страница 30: ...30 N2820 1A User s Guide 2 Making Measurements Quick Start Figure 10 R Sense Entry Box...
Страница 74: ...74 N2820 1A User s Guide 4 Probe Calibration To Calibrate the Probe on InfiniiVision Oscilloscopes...
Страница 84: ...84 N2820 1A User s Guide 5 Characteristics and Specifications Dimensions...
Страница 94: ...94 N2820 1A User s Guide 6 Performance Data Plots Input Load Model...
Страница 106: ...106 N2820 1A User s Guide 7 Performance Verification Performance Test Record...
Страница 109: ...N2820 1A User s Guide 109...
Страница 110: ...This information is subject to change without notice Keysight Technologies 2018 March 2018 N2820 97006 www keysight com...