Probe Calibration
4
To Calibrate the Probe on Infiniium Oscilloscopes
N2820/1A User’s Guide
67
4
Connect the equipment as shown in
the following requirements:
• Connect the probe to the scope channel that you will use during
probing. Use any other available scope channel for the BNC
cable that is needed for connection to the Pomona 3788.
• Connect the Pomona 3788 to two BNC cables using a BNC tee.
Then connect one of these BNC cables to a scope channel and
the other BNC cable to the Aux Output BNC connector located
on the Oscilloscope’s rear-panel (as shown in
• On N2820A probes, do not connect the secondary cable at this
time.
• Connect both the N2825A’s sensor leads (red and black wires) to
the black test clip of the Pomona 3788 as shown in
.
Figure 30
Example of Initial Calibration Setup
N2820A Probe
N2825A Probe Head
Black Test Clip of
Pomona 3788
BNC Tee
BNC
Cable
BNC
Cable
Red Test Clip of
Pomona 3788
Pomona 3788
Содержание N2820A
Страница 1: ...Keysight N2820 1A High Sensitivity Current Probes User s Guide...
Страница 30: ...30 N2820 1A User s Guide 2 Making Measurements Quick Start Figure 10 R Sense Entry Box...
Страница 74: ...74 N2820 1A User s Guide 4 Probe Calibration To Calibrate the Probe on InfiniiVision Oscilloscopes...
Страница 84: ...84 N2820 1A User s Guide 5 Characteristics and Specifications Dimensions...
Страница 94: ...94 N2820 1A User s Guide 6 Performance Data Plots Input Load Model...
Страница 106: ...106 N2820 1A User s Guide 7 Performance Verification Performance Test Record...
Страница 109: ...N2820 1A User s Guide 109...
Страница 110: ...This information is subject to change without notice Keysight Technologies 2018 March 2018 N2820 97006 www keysight com...