212
Keysight B2980B User’s Guide, Edition 1
Function Details
Null, Offset Cancel
Resistance Measurement (B2985B/B2987B)
If the Null key is pressed when the NULL indicator is turned off, the current
resistance measurement value is set as the offset value. Just after that, the
resistance measurement is performed once, and the value that subtracted the
offset value from the actual measured value is displayed as the measurement
result. And also the NULL indicator is turned on.
If the Null key is pressed when the NULL indicator is tuned on, the null function is
disabled and the NULL indicator is turned off. Just after that, the resistance
measurement is performed once, and the actual measurement value is displayed.