Keysight B2980B User’s Guide, Edition 1
241
Function Details
Trigger System
Table 5-7
Trigger Output Timing and Setup Parameters
For using a digital I/O pin, it is necessary to set function of the pin, output trigger
polarity, type, pulse width, and the Output Trigger Timing parameter. See
Configuration Dialog Box” on page 194
and
“Using Digital I/O and Trigger In/Out”
.
For using the Trigger Out connector, it is necessary to set output trigger polarity,
type, pulse width, and so on. See
“BNC Configuration Dialog Box” on page 195
.
Trigger output timing
Layer
Action
Trigger
Output
Output
Trigger
Timing
Starting ARM loop
ARM
TRANS. for
transient
action or ACQ.
for acquire
action
ON
BEFORE
Ending ARM loop
ARM
ON
AFTER
Starting TRIGGER loop
TRIGGER
ON
BEFORE
Ending TRIGGER loop
TRIGGER
ON
AFTER
Starting Transient device action
ACTION
TRANS.
ON
BEFORE
Ending Transient device action
ACTION
TRANS.
ON
AFTER
Starting Acquire device action
ACTION
ACQ.
ON
BEFORE
Ending Acquire device action
ACTION
ACQ.
ON
AFTER
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...