Keysight B2980B User’s Guide, Edition 1
99
How to Perform Measurement
Voltage Measurement
The guarded voltage measurement is effective when the guarded voltmeter
indicator turns on. In this setup, the same potential voltage appears at the center
conductor (Input) and the inner shield through a buffer amplifier (
1) as shown in
. It is effective for performing voltage measurement accurately.
See
Measure Volts
Press the RANGE function key to change the ranging mode, auto or fixed. Where,
The softkey label shows the present measurement range setting, AUTO or fixed
range value.
Measurement range can be set by using the following navigation keys.
Range +
Sets the fixed range mode and changes the range up.
[
home
]
Sets the AUTO range and the Normal auto aperture.
Range
Sets the fixed range mode and changes the range down.
See
“Measurement Range” on page 89
for more information.
Aperture Time
Press the SPEED function key to change the aperture mode, auto or manual.
Where, the softkey label shows the present aperture time setting.
Aperture time can be set by using the following navigation keys.
Coarse Res
Increases measurement speed (decreases aperture time).
[
home
]
Sets the Normal auto aperture and the AUTO range.
Fine Res
Decreases measurement speed (increases aperture time).
See
for more information.
Filter
Measurement filter function can be enabled/disabled by using the Filter key.
If the FILT indicator is turned off, pressing the Filter key opens the Measure Filter
dialog box. Pressing OK after setting up the dialog box enables the filter function
and turns the FILT indicator on.
If the FILT indicator is turned on, pressing the Filter key disables the function and
turns the FILT indicator off.
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...