126
Keysight B2980B User’s Guide, Edition 1
How to Perform Measurement
Measurement Considerations
Measurement Considerations
This section describes some considerations for performing accurate and reliable
measurements.
•
•
“Leakage Current on Connection Parts”
•
•
•
•
•
•
•
Insulating Material
The need to use high-resistance insulation materials in connection parts, such as
cables, adapters, and others, is essential to ensure the reliability of ultra-low
current measurements. Poor insulation will allow greater leakage current.
Leakage Current on Connection Parts
Contaminants such as moisture or ionic chemicals can cause electrochemical
effects that degrade insulation resistance. In some cases, ionic chemicals create a
battery effect that sources offset current. This effect is not stable and will be a
major obstacle to obtaining reliable low-current measurements. It is important to
keep the surface of insulation material clean.
Humidity and Temperature
Water and water vapor can easily cause electrochemical effects. It is effective to
maintain constant low humidity in the measurement environment in order to
prevent the occurrence of an electrochemical effect. Temperature changes can
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...