186
Keysight B2980B User’s Guide, Edition 1
Front Panel Reference
Result Key Group
Length
Data length
The limit test data contains the following information.
(aaaaa) BIN: bb DATA: +c.dd
(aaaaa)
Data index
aaaaa
BIN:
Bin number
bb
(
01
to
12
)
The bin numbers 1 to 12 are associated with the index numbers
1 to 12. See
“Limit Test Setup Dialog Box” on page 180
.
If the limit test data is out of the bins,
00
is set for the GRADING
mode and
15
is set for the SORTING mode.
DATA:
Limit test data
+c.dd
Trace Statistical Result Dialog Box
This dialog box provides the following GUI for displaying the trace statistical
result. The result data is displayed in the Mean, Std. Dev., Min., and Max. fields.
Feed
Type of data to display, SENSE, MATH, or LIMIT
SENSE: Measurement result data
MATH: Calculation result data of math expression
LIMIT: Limit test data
Element
Element type of the measurement result data to display. This is
available only when the Feed parameter is set to SENSE.
AMPS: Current measurement data
COUL.: (B2985B/B2987B) Charge measurement data
VOLTS: (B2985B/B2987B) Voltage measurement data
OHMS: (B2985B/B2987B) Resistance data
Length
Data length
Mean
Mean value
Std. Dev.
Standard deviation
Min.
Minimum value
Max.
Maximum value
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...