How to Perform Measurement
Resistance Measurement
Voltage Source
The B2985B/B2987B supports the voltage source capability shown in
.
Table 3-4
Voltage Output Range, Value, Resolution, and Maximum Current
Figure 3-16
Simplified Circuit Diagram of Voltage Source
Procedure 1
You can apply DC voltage as follows.
Step 1. Press the
View
key and select the Meter View function key to display the Meter
view.
Step 2. Use the Spot Source Range field to set the output range you want to use. See
“DC Voltage Output Range” on page 115
for more information.
Step 3. Press the
More... 1 of 3
,
More... 2 of 3
, and
Source
assist keys to move the
field pointer onto the Voltage Source field.
Step 4. Set the output voltage value to the Voltage Source field. Then use the rotary
knob and/or the assist key.
Step 5. Press the Voltage Source
On/Off
switch to enable the output. This turns the
switch green. And the voltage source starts output.
Step 6. Press the Voltage Source
On/Off
switch to disable the output. This turns off the
switch light.
Procedure 2
You can apply sweep voltage as follows.
Range value
Voltage output value
Display resolution Maximum current
20 V
0
|V|
21 V
700
V
20 mA
1000 V
0
V
1050 V
35 mV
1 mA
1000 V
1050 V
V
0
High
Low
On
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...