96
Keysight B2980B User’s Guide, Edition 1
How to Perform Measurement
Voltage Measurement
Figure 3-8
Unguarded Voltage Measurement Connection
Instead of connecting the inner shield (Common) of Voltmeter, a banana to
alligator clip cable can be used for connecting Common to low terminal of voltage
under test.
For floating voltmeter, do not connect any cable between Common and chassis
“Common Terminal Connection” on page 92
for more information.
Procedure
You can perform the voltage measurement as follows.
Step 1. Press the
View
key and the Meter View function key to display the Meter view.
Step 2. Press the
VOLTS (V)
assist key to set the voltage measurement mode.
Step 3. Use the navigation keys or the RANGE function key to set the measurement
range you want to use. See
for more information.
Step 4. Use the navigation keys or the SPEED function key to set the measurement
speed (aperture time) you desire. See
for more
information.
Vs
Chassis ground
Triaxial cable
Common is connected to chassis ground by using Banana-to-lug cable.
Triaxial bulkhead connector
V
Vs
Common
Chassis
Ground
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...