104
Keysight B2980B User’s Guide, Edition 1
How to Perform Measurement
Resistance Measurement
Pour fixer l'adaptateur N1413A/N1414A, enfoncez-le dans les connecteurs
B2985B/B2987B.
Poussez l'adaptateur fermement jusqu'à ce qu'il se verrouille en position (espace
< 1 mm).
For using the N1413A/N1414A adapter, set the Low Terminal State to FLOATING.
Voltage Source Low is connected to the chassis ground through the adapter.
If you use the N1413A/N1414A adapter, you do not need to care about setting of
the Voltmeter connector’s inner shield. It is opened at the adapter’s input.
Figure 3-12
Floating Device Measurement
Voltage Source Low terminal is internally connected to or disconnected from the
. See
“Low Terminal State” on page 110
for more information.
Input
Triaxial cable
Triaxial bulkhead connector
High voltage test lead
R
X
High
Common is connected to chassis ground by using Banana-to-lug cable.
Shield
Input
R
X
High
Shield
A
V
S
Voltage Source Low Terminal State:
CCOM
Common
Banana-to-lug
cable
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...