Keysight B2980B User’s Guide, Edition 1
153
Front Panel Reference
Display and Assist/Function Keys
Source Function Sub-panel (B2985B/B2987B)
Use this sub-panel to set up the voltage source in the B2985B/B2987B as a sweep
source or a square waveform generator.
To select the source function, use the assist keys displayed when the field pointer
is in the EDIT (green) status on the VS Function field.
•
LINEAR SINGLE: Sweep from start to stop in a linear incremental step
•
LINEAR DOUBLE: Sweep from start to stop to start in a linear incremental step
•
LIST: Sweep of values defined in the List sweep setup list. See
.
•
ARB SQUARE: Square waveform output
•
OFF: The voltage source works as the DC constant voltage source.
The following setup parameters are available for LINEAR SINGLE, LINEAR
DOUBLE, or LIST.
Start
Sets the sweep start value.
Stop
Sets the sweep stop value.
Points
Sets the number of sweep steps.
Step
Sets the sweep step value (incremental step value). Not
available for the LIST sweep operation.
The following setup parameters are available for ARB SQUARE.
Start
Sets the base voltage of square waveform.
Peak
Sets the peak voltage of square waveform.
Delay
Sets a period of base output before peak.
Peak Time
Sets a period of peak output.
End Time
Sets a period of base output after peak.
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...