252
Keysight B2980B User’s Guide, Edition 1
Function Details
Setup Integrity Checker
Execution Examples
Example 1
Shielding Effect for the 1.5 m Triaxial Cable
This example compares the case with connecting the Triaxial cable only (the left
side in the figure) and the case with connecting the Triaxial cable and the shield
case that is connected to the chassis (the right side in the figure, the core wire and
inner shield of Triaxial cable are open). In the comparison, the noise checking was
performed 5 times and the average of the results is calculated in both cases.
The average of Variance Ratio seems to be improved by about 2/3 due to the
shielding effect.
• Executes noise checking 5 times at PLC=1 and
20 pA range
• 1.5 m Triaxial Cable
• Cable end is OPEN (no shielding)
• Results
• Average of Target = 1.42E -15
• Average of Variance Ratio = 3.12
Triaxial Cable
Triaxial Cable
=
x 2/3
• Executes noise checking 5 times at PLC=1 and
20 pA range
•
• Cable end is OPEN (shielding)
• Results
• Average of Target 1.22E -15
• Average of Variance Ratio =1.96
1.5 m Triaxial Cable
Shielding
Содержание B2980B Series
Страница 13: ...Keysight B2985B B2987B High Low 1050 Vdc 3 Interlock High Low On Off On Off HV Interlock...
Страница 16: ......
Страница 24: ......
Страница 52: ...52 Keysight B2980B User s Guide Edition 1 Introduction Options...
Страница 80: ...80 Keysight B2980B User s Guide Edition 1 Installation Using Digital I O and Trigger In Out...
Страница 130: ...130 Keysight B2980B User s Guide Edition 1 How to Perform Measurement Measurement Considerations...
Страница 202: ...202 Keysight B2980B User s Guide Edition 1 Front Panel Reference System Key Group...
Страница 264: ...264 Keysight B2980B User s Guide Edition 1 Function Details Initial Settings...