4-52 Keysight N5247B Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
3.
Connect the spectrum analyzer to the open bypass switch connector. Set
the spectrum analyzer to measure a signal at 800 MHz.
4.
If the 800 MHz signal is not present, replace the mechanical switch. Refer
“Removing and Replacing the A50–A53 Bypass Switches and the A54
5.
If the 800 MHz signal is present, replace the associated reference coupler.
Refer to
“Removing and Replacing the A29 –A32 Reference Couplers and
Reference Coupler Mounting Brackets” on page 7-43
.
Checking the A70–A74 Bias Combiners
Refer to
“Checking the A70/A75 Low Frequency Extension (LFE) Board and
the Bias Combiners” on page 4-47
Checking the Receiver Group
Equipment Used for These Tests
For all of the following checks, refer to the block diagrams at the end of this
chapter and to any of the following that are appropriate:
—
“2-Port Configurations, Serial Number Prefixes <6021” on page 6-31
—
“4-Port Configuration, Serial Number Prefixes <6021” on page 6-84
—
“2-Port Configurations, Serial Number Prefixes
—
“4-Port Configurations, Serial Number Prefix
Getting Ready to Test
Before checking the assemblies, you must open the analyzer.
1.
Turn off the analyzer power.
2.
Unplug the power to the analyzer and disconnect all front and rear panel
connections except installed jumpers.
Equipment Type
Model or
Part Number
Alternate Model or Part
Number
Spectrum analyzer
8565E
856xE
a
a.
Refer to
. Must be capable of measuring a signal at 7.61 MHz
(analyzers with DSP version 4.0), or 7.44 MHz (analyzers with DSP version 5.0), and 1 GHz.
Use an antistatic work surface and wrist strap to reduce the chance of
electrostatic discharge for all of the procedures in this chapter.
Содержание N5247B
Страница 10: ...10 Keysight N5247B Service Guide Contents ...
Страница 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Страница 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Страница 227: ...Keysight N5247B Service Guide 5 57 Theory of Operation Receiver Group Operation Figure 5 18 4 Port Receiver Group ...
Страница 228: ...5 58 Keysight N5247B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 19 2 Port Receiver Group ...
Страница 238: ...5 68 Keysight N5247B Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 243: ...Keysight N5247B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure ...
Страница 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Страница 381: ...Keysight N5247B Service Guide 6 143 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 548: ...7 104 Keysight N5247B Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 565: ...Installation Note Xxxxx xxxxx 3 ...