4-44 Keysight N5247B Service Guide
Troubleshooting
Measurement System Troubleshooting
4-
4.
Set the network analyzer for an 800 MHz CW frequency and observe the
spectrum analyzer measurement. An 800 MHz signal should be present.
5.
Refer to
. If the observed problem was
frequency banded rather than broadband related, set the analyzer
frequency to the center of the problem band. The spectrum analyzer
should measure a signal above the network analyzer setting. For analyzers
with DSP version 4.0, the signal is 7.61 MHz above the network analyzer
setting. For analyzers with DSP version 5.0, the signal is 7.44 MHz above
the network analyzer setting.
6.
If the signal is not present but the 50 MHz reference signal from
the A14 50 MHz Reference Outputs”
is present, replace the faulty
“Removing and Replacing the A4–A17 Boards”
7.
If the signal is present, reconnect cables W1 and W2, and then continue
with
“Checking the A5 and A10 Source Outputs”
.
Checking the A5 and A10 Source Outputs
1.
Refer to the block diagram at the end of this chapter and to
All Cables—All Options, S/N Prefixes <6021” on page 6-21
. Locate the
cables W3 and W4 at the A5 source 1 board or cables W7 and W8 at the
A10 source 2 board. (W7, W8, and A10 are only available in 4-port models
and 2-port models with Option 224.)
2.
Disconnect cables W3 and W4 or cables W7 and W8, dependent on which
source board is to be checked, at the A7, A8, A12, or A13 40 GHz Doubler
board.
3.
Connect the spectrum analyzer to the open connector at the end of the
cable that connects to the source board to be checked.
4.
Set the network analyzer for an 800 MHz CW frequency and observe the
spectrum analyzer measurement. An 800 MHz signal should be present.
5.
Refer to
. If the observed problem was
frequency banded rather than broadband related, set the analyzer
frequency to the center of the problem band. The spectrum analyzer
should measure a signal above the network analyzer setting. For analyzers
with DSP version 4.0, the signal is 7.61 MHz above the network analyzer
setting. For analyzers with DSP version 5.0, the signal is 7.44 MHz above
the network analyzer setting.
6.
If the signal is not present but the signals from
are present, replace the appropriate source
board. Refer to
“Removing and Replacing the A4–A17 Boards” on
.
7.
If the signal is present, reconnect all cables, and then continue with
“Checking the A7, A8, A12, and A13 40 GHz Doubler Outputs”
.
Содержание N5247B
Страница 10: ...10 Keysight N5247B Service Guide Contents ...
Страница 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Страница 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Страница 227: ...Keysight N5247B Service Guide 5 57 Theory of Operation Receiver Group Operation Figure 5 18 4 Port Receiver Group ...
Страница 228: ...5 58 Keysight N5247B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 19 2 Port Receiver Group ...
Страница 238: ...5 68 Keysight N5247B Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 243: ...Keysight N5247B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure ...
Страница 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Страница 381: ...Keysight N5247B Service Guide 6 143 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 548: ...7 104 Keysight N5247B Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 565: ...Installation Note Xxxxx xxxxx 3 ...