3-54 Keysight N5247B Service Guide
Tests and Adjustments
Adjustments
3-
Adjustments
These adjustments are firmware-driven tests that are used to fine-tune your
analyzer.
If multiple adjustments are to be performed, perform them in the order listed.
—
“10 MHz Frequency Reference Adjustment” on page 3-55
—
“Synthesizer Bandwidth Adjustment” on page 3-56
—
“Source Adjustment” on page 3-56
—
“IF Gain Adjustment” on page 3-58
—
“Receiver Characterization” on page 3-58
—
“Receiver Adjustment” on page 3-58
—
“EE Default Adjustment” on page 3-60
—
“IF Response Adjustment (Available Only with Option S93093/4A/B and or
S93900/1/2/4/5/7A/B Installed)” on page 3-61
—
“LFE Receiver Adjustment (Option 425)” on page 3-61
—
“Noise Figure Adjustment (Available with Option 029 with S93029A/B
These adjustments are described on the following pages.
Touchscreen Adjustment and Verification
The touchscreen adjustment (calibration) is used to correct the touchscreen
accuracy when you replace your front panel, or your SSD. Refer to
The touchscreen verification can be run to verify the touchscreen has been
correctly calibrated. Refer to
“Verification Procedure” on page 3-55
.
Calibration Procedure
Step 1.
Verify your instrument is powered up.
Step 2.
Press
Instrument
>
Display
>
Display Setup
>
Calibrate Touchscreen...
Step 3.
In the Table PC Setting window that opens, press
Calibrate...
For best results, use a touchscreen soft touch pen for the following
adjustments.
If your touchscreen is not behaving as expected, you can reset your display using the Table PC
Setting window and re-calibrate the touch settings using the
below.
Содержание N5247B
Страница 10: ...10 Keysight N5247B Service Guide Contents ...
Страница 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Страница 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Страница 227: ...Keysight N5247B Service Guide 5 57 Theory of Operation Receiver Group Operation Figure 5 18 4 Port Receiver Group ...
Страница 228: ...5 58 Keysight N5247B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 19 2 Port Receiver Group ...
Страница 238: ...5 68 Keysight N5247B Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 243: ...Keysight N5247B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure ...
Страница 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Страница 381: ...Keysight N5247B Service Guide 6 143 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 548: ...7 104 Keysight N5247B Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 565: ...Installation Note Xxxxx xxxxx 3 ...