Keysight N5247B Service Guide 4-41
Troubleshooting
Measurement System Troubleshooting
setting. For analyzers with DSP version 4.0, the signal is 7.61 MHz above
the network analyzer setting. For analyzers with DSP version 5.0, the
signal is 7.44 MHz above the network analyzer setting
6.
If the signal is not present but the signal is present, from
A15 13.5 GHz LO Synthesizer Output”
(S/N prefixes
<
6021) or
the A15 Direct Digital Synthesizer (DDS) Assembly Outputs”
(LO Output,
prefixes
≥
6021), replace the A25 HMA 26.5. Refer to
Replacing the A25 HMA26.5” on page 7-36
7.
If the signal is present, reconnect cable W52 or W80 and then:
—
for 2-port models, continue checking with
.
—
for 4-port models, continue checking with
Verify the Model/Version of HMA26.5 Installed
The installed components in your PNA model may be either using the legacy
HMA26.5 part number 5087-7765 or if your PNA is using the newer HMA26.5
part number N5240-60101.
Legacy HMA26.5 Use the Following Components:
— A22 splitter 5087-7139
— W42 N5222-20009
— W43 N5222-20007
— W44 N5222-20008
The new N5240-60101 HMA26.5 has the splitter integrated into the assembly
and uses an N5222-20126 cable. Refer to
.
See also
Chapter 6, “Replaceable Parts.”
and the “Bottom Assemblies and
Cables” section for your configuration and serial number prefix.
Содержание N5247B
Страница 10: ...10 Keysight N5247B Service Guide Contents ...
Страница 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Страница 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Страница 227: ...Keysight N5247B Service Guide 5 57 Theory of Operation Receiver Group Operation Figure 5 18 4 Port Receiver Group ...
Страница 228: ...5 58 Keysight N5247B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 19 2 Port Receiver Group ...
Страница 238: ...5 68 Keysight N5247B Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 243: ...Keysight N5247B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure ...
Страница 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Страница 381: ...Keysight N5247B Service Guide 6 143 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 548: ...7 104 Keysight N5247B Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 565: ...Installation Note Xxxxx xxxxx 3 ...