Keysight N5247B Service Guide 4-35
Troubleshooting
Measurement System Troubleshooting
Checking the Source Group
Source Group Tests
Before checking the source group assemblies, you must open the analyzer.
1.
Turn off the analyzer power.
2.
Unplug the power to the analyzer and disconnect all front and rear panel
connections except installed jumpers.
3.
Remove the outer and inner covers from the analyzer. Refer to
.
4.
With the covers off, plug in the analyzer and turn on the power.
Frequency Banded vs. Broadband Failure
There are two main types of failures that are related to the source group.
Frequency banded failures are indicated by all receiver traces having partial
dropouts across the frequency range. Broadband failures are indicated by all
receiver traces being in the noise floor.
RF Signal Troubleshooting
Check the output power of the A, B, C, and D signals:
Equipment Used for This Check
Use an antistatic work surface and wrist strap to reduce the chance of
electrostatic discharge for all of the procedures in this chapter.
Procedures described in this document are performed with power
supplied to the product while protective covers are removed. Energy
available at many points may, if contacted, result in personal injury.
Do not operate the analyzer with the outer cover removed for more than
30 minutes, as this could cause the analyzer to overheat which could
result in costly damage.
IMPORTANT!
In 2020, the PNA added direct digital synthesis (DDS)
capability to all serial number prefix
≥
6021 instruments. The effect of
installing this DDS assembly is that the A4, A10, A11
a
, A13 boards are
integrated into a single A11
direct digital synthesizer assembly, and the
, and A13 boards do
not
apply to serial number
prefix instruments
≥
6021.
a. If your PNA has a serial prefix
≥
6021
or has been upgraded with a version 7 synthesizer assembly,
then A15 is a direct
digital synthesizer (DDS) assembly. If your PNA has a serial prefix
<
6021, then A11 is a 13.5 GHz synthesizer board.
Содержание N5247B
Страница 10: ...10 Keysight N5247B Service Guide Contents ...
Страница 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Страница 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Страница 227: ...Keysight N5247B Service Guide 5 57 Theory of Operation Receiver Group Operation Figure 5 18 4 Port Receiver Group ...
Страница 228: ...5 58 Keysight N5247B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 19 2 Port Receiver Group ...
Страница 238: ...5 68 Keysight N5247B Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 243: ...Keysight N5247B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure ...
Страница 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Страница 381: ...Keysight N5247B Service Guide 6 143 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 548: ...7 104 Keysight N5247B Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 565: ...Installation Note Xxxxx xxxxx 3 ...