Keysight N5247B Service Guide 3-41
Tests and Adjustments
Performance Tests
4.
A2 is set to 20 dB and the analyzer’s receiver power level is measured. This
is the analyzer reference reading, Par.
5.
A1 is changed to 5 dB and 10 dB. At each point, delta power levels are
read on the power meter, Pmd, and the analyzer, Pad.
6.
The power linearity error at each point is calculated as Pe = (Pmr – Pmd) –
(Par – Pad).
7.
A2 is set to 30 dB, A1 is set to 0 dB, and the analyzer’s source power is
adjusted until the receiver power level is exactly the same as it was before
the attenuators were switched.
8.
New power meter and analyzer receiver reference readings are recorded.
The process is repeated until the total attenuation reaches 120 dB.
9.
The process is reset to the 20 dB reference level and it is run in reverse
until the total attenuation reaches 0 dB.
If the Analyzer Fails this Test:
— If the analyzer fails this test, rerun the test.
— If the analyzer fails this test repeatedly, replace the A23 mixer brick for a
Port 1 or Port 2 failure or the A24 mixer brick for a Port 3 or Port 4 failure,
then repeat this test. Refer to
“Removing and Replacing the A27 and A28
— If the analyzer still fails this test, replace the A12 SPAM board and repeat
this test. Refer to
“Removing and Replacing the A4–A17 Boards” on
.
Dynamic Accuracy Test – Version 2
This description applies to all N522xB, N5231B/32B/34B/35B/39B, and
N5247B instruments. It also applies to N5241B/42B/44B/45B instruments
with serial numbers greater than xx5240xxxx. It also applies to
N5241B/42B/44B/45B instruments with serial numbers less than xx5240xxxx.
For all other instruments, see test
“Dynamic Accuracy Test – Version 1” on
“Dynamic Accuracy at Low Frequency (Available with LFE
Option 205/220/420/425)” on page 3-43
.
Function of the Test: To measure the relative power linearity of the analyzer’s
receivers.
Specification Tested: Test Port Input–Dynamic Accuracy
Equipment Used:
— U3020AD01 dynamic accuracy test set
— Signal generator
— Power meter
— Power sensor
Содержание N5247B
Страница 10: ...10 Keysight N5247B Service Guide Contents ...
Страница 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Страница 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Страница 227: ...Keysight N5247B Service Guide 5 57 Theory of Operation Receiver Group Operation Figure 5 18 4 Port Receiver Group ...
Страница 228: ...5 58 Keysight N5247B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 19 2 Port Receiver Group ...
Страница 238: ...5 68 Keysight N5247B Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 243: ...Keysight N5247B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure ...
Страница 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Страница 381: ...Keysight N5247B Service Guide 6 143 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 548: ...7 104 Keysight N5247B Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 565: ...Installation Note Xxxxx xxxxx 3 ...