![Keysight Technologies N5247B Скачать руководство пользователя страница 63](http://html.mh-extra.com/html/keysight-technologies/n5247b/n5247b_service-manual_1962042063.webp)
Keysight N5247B Service Guide 3-33
Tests and Adjustments
Performance Tests
Description of the Test:
1.
The analyzer is Preset.
2.
The analyzer is set up for a CW reflection measurement on the test port to
be measured.
3.
A power sensor is connected to the test port, using any necessary
adapters.
4.
The analyzer frequency is set to the desired value.
5.
The power meter correction table is set to the same frequency.
6.
The output power is measured, and the value is compared to the Preset
setting.
7.
This process is repeated at hundreds of frequencies across the analyzer’s
full range. The difference between the measured power and the output
setting must fall within the specified accuracy range at all points for the
test to pass.
If the Analyzer Fails this Test:
— Perform the
“Source Adjustment” on page 3-56
and repeat this test.
— If the analyzer still fails this test, troubleshoot the source section of the
analyzer and then repeat this test. Refer to
“Checking the Source Group” on
.
Source Maximum Power Output Test
Function of the Test: To confirm the maximum source output power of your
network analyzer over its full frequency range.
Specification Tested: Test Port Output–Maximum Leveled Power
Equipment Used:
— Power meter
— Power sensors
— Any necessary adapters
Description of the Test:
1.
The analyzer is Preset.
2.
The analyzer is set up for a CW reflection measurement on the test port to
be measured.
3.
A power sensor is connected to the test port, using any necessary
adapters.
4.
The analyzer frequency is set to the desired value.
5.
The power meter correction table is set to the same frequency.
Содержание N5247B
Страница 10: ...10 Keysight N5247B Service Guide Contents ...
Страница 96: ...3 66 Keysight N5247B Service Guide Tests and Adjustments Adjustments 3 ...
Страница 157: ...Keysight N5247B Service Guide 4 61 Troubleshooting This page left intentionally blank ...
Страница 227: ...Keysight N5247B Service Guide 5 57 Theory of Operation Receiver Group Operation Figure 5 18 4 Port Receiver Group ...
Страница 228: ...5 58 Keysight N5247B Service Guide Theory of Operation Receiver Group Operation 5 Figure 5 19 2 Port Receiver Group ...
Страница 238: ...5 68 Keysight N5247B Service Guide Theory of Operation Noise Measurement Group Operation Option 029 5 ...
Страница 243: ...Keysight N5247B Service Guide 6 5 Replaceable Parts Rebuilt Exchange Assemblies Figure 6 1 Module Exchange Procedure ...
Страница 268: ...6 30 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 270: ...6 32 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 373: ...Keysight N5247B Service Guide 6 135 Replaceable Parts Replaceable Parts Listings ...
Страница 381: ...Keysight N5247B Service Guide 6 143 Replaceable Parts Replaceable Parts Listings This page intentionally left blank ...
Страница 426: ...6 188 Keysight N5247B Service Guide Replaceable Parts Replaceable Parts Listings 6 ...
Страница 548: ...7 104 Keysight N5247B Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 565: ...Installation Note Xxxxx xxxxx 3 ...