7.2
Weighing units switch-over
Using the
Unit
-key you can switch over the display between
kg
,
lb
and
N
. The
X
marks the active setting.
7.3 Taring
•
Switch on the balance using the
ON/OFF/TARE
-key and wait
for the
„0“
display. Suspend the tare load and press the
ON/OFF/TARE
-key.
•
The balance display jumps to „
0.0
“. The weight of the tare
load is internally stored. The material to be weighed can now
be filled in the tare vessel and the weighing value can be
read.
7.4 Functions
With help of the
Hold
-key the following functions can be acti-
vated:
Setting Function
H1
Data-Hold function 1
Weighing value frozen for 5 sec. after press-
ing the
Hold
-key
H2
Data-Hold function 2
The weighing value will be frozen after press-
ing the
Hold
-key until another key is actuated
H3
Data-Hold function 3
The weighing value is automatically frozen
for 5 sec.
H4
Data-Hold function 4
The weighing value is automatically frozen
after reaching a stable value until a key will
be actuated
H5
Animal weighing function
H6
Peak value function
7.4.1 Data-Hold
function
With the Data-Hold function the weighing values in the display
can be fixly displayed for a longer time. According to the using
range you can select between 4 settings (H1 – H4).
Ö
Switch on the balance, keep the HOLD-key pressed until
the current setting „Hx“ (H1 – H6) appears.
Ö
Actuate
the
ON/OFF/TARE
-key until the desired setting
(H1-H4) appears.
Ö
Use
the
HOLD
-key to confirm your selection
Ö
Suspend the material to be weighed
Ö
The weighing value is displayed depending on your set-
tings (H1 – H4) (see table chap. 7.4), symbolised by the
[
S
] top left.
7.4.2
Animal weighing function
This function is suitable for busy weighing procedures. The
result is a mean value formed by 16 weighing values which is
found out within 3 seconds.
Ö
Switch on the balance, keep the
HOLD
-key pressed until
the current setting „Hx“ (H1 – H6) appears.
Ö
Press
the
ON/OFF/TARE
-key until „
H5
“ appears.
Ö
Use the
HOLD
-key to confirm your selection
Ö
Suspend the material to be weighed
Ö
Press
the
HOLD
-key, the display counts reverse from
3 -1. The calculated mean value is indicated, symbolised
by the [
S
] top left.
Ö
For further measurements press the
ON/OFF/TARE
-key
before
7.4.3
Peak value function
This function displays the highest load value (peak value) of a
weighing.
Measuring frequency: 200ms
Caution:
Never exceed the maximum permitted load of the
peak value on the balance (!!Danger of breaking!!).
Ö
Switch on the balance, keep the
HOLD
-key pressed until
the current setting „Hx“ (H1 – H6) appears.
Ö
Press
the
ON/OFF/TARE
-key until „
H6
“ appears.
Ö
Use the
HOLD
-key to confirm your selection
Ö
Suspend the material to be weighed
Ö
The peak value appears for a short time, symbolised by
the [
S
] top left. The balance returns automatically to zero
and is ready for further measurement.
8
Service, maintenance, disposal
8.1 Cleaning
Please do not use aggressive cleaning agents (solvents or
similar agents), but a cloth dampened with mild soap suds.
Ensure that no liquid penetrates into the device and wipe with a
dry soft cloth.
8.2 Service,
maintenance
The appliance may only be opened by trained service techni-
cians who are authorized by KERN.
8.3 Disposal
Disposal of packaging and appliance must be carried out by
operator according to valid national or regional law of the loca-
tion where the appliance is used.
9 Instant
help
In case of a fault in the program sequence, the balance should
be shortly switched off. The weighing process must then be
restarted from the beginning.
Fault
Possible cause
•
The balance is not switched on.
Display weight
does not glow.
•
Batteries are inserted incorrectly or empty
•
No batteries inserted.
•
Draught/air movement
Displayed
weight is per-
manently
changing
•
Vibrations on the supension of the balance
•
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
Weighing
result obvi-
ously incorrect
•
The display of the balance is not at zero
•
Great fluctuations in temperature.
•
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
Should other error messages occur, switch balance off and
then on again. If the error message remains inform manufac-
turer.