Models 2510 and 2510-AT
Specifications
A-3
Accuracy calculations
The information below discusses how to calculate accuracy for both TEC measurement and
thermal feedback element specifications.
Accuracy specifications are stated as follows:
Accuracy = ±(% of r offset)
As an example of how to calculate the actual limits, assume an output voltage of 5V. You
can compute the limits from one-year operating voltage accuracy specifications as follows:
Accuracy =
±(% of r offset)
±[(0.1% X 5V) + 4mV]
±(5mV + 4mV)
±9mV
Thus, the actual operating voltage range is: 5V± 9mV, or from 4.991 to 5.009V.
2510-902-01 (E - Mar 2010)(Service).book Page 3 Tuesday, March 16, 2010 2:33 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...