Models 2510 and 2510-AT
Troubleshooting
4-17
Analog circuitry checks
Table 4-4 summarizes analog circuitry checks.
NOTE
All measurements are with the V function selected, +5V output voltage, 5A current
limit, and 1
Ω
and 100
Ω
resistors connected to the OUTPUT and INPUT terminals
respectively. Turn OUTPUT ON while measuring.
Internal fuse replacement
WARNING
Disconnect the line cord and all cables and test leads from the instrument
before replacing internal fuses.
CAUTION
Do not install a fuse with a higher current rating than specified, or instru-
ment damage may occur.
Power supply module fuse replacement
An internal fuse protects the power supply module from over-current conditions. Replace
this fuse as follows:
1.
Turn off the power, and disconnect the line cord and all other test leads and cables from
the instrument.
2.
Remove the case cover and mother board, as covered in Section 5.
3.
Locate the fuse on the power supply module circuit board. The module is mounted on
the chassis bottom.
4.
Replace the fuse with one with the same current and voltage rating marked on the
power supply module circuit board.
Table 4-4
Analog circuitry checks
Step Item/component
1
Required condition
Remarks
1
2
3
4
5
6
7
8
9
TP103
1
TP104
1
TP108
2
TP109
2
TP110
2
TP114
2
TP118
2
TP119
2
TP120
2
+10.4V
+4V
+6.3V
-12.5V
+12.5V
0V
Pulse train
0.5V
5V
H drive signal
H drive signal
A/D reference signal
-14V nominal supply
+14V nominal supply
Sensor feedback signal
Mux output to A/D converter
OUTPUT current sense signal
OUTPUT voltage sense signal
1. Measured with respect to F common (TP121).
2. Measured with respect to A2 common (TP122).
2510-902-01 (E - Mar 2010)(Service).book Page 17 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...