Models 2510 and 2510-AT
Calibration
2-25
16.
Send this command to select the 100k
Ω
resistance range:
:SENS:TEMP:THER:RANG 1e5
17.
Send the following command with the characterized 100k
Ω
resistance value as the
parameter to calibrate the 100k
Ω
range:
:CAL:PROT:SENS:TEMP <Resistance_Value>
18.
Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for
thermal equilibrium, then send this command:
:CAL:PROT:SENS:TEMP 0
19.
Connect the 100
Ω
resistor to the INPUT terminals (Figure 2-3).
20.
Send the following command to select the RTD type temperature sensor:
:SENS:TEMP:TRAN RTD
21.
Send this command to select the 100¾ range:
:SENS:TEMP:RTD 100
22.
Send the following command with the characterized 100
Ω
resistance value as the
parameter to calibrate the RTD measurement function:
:CAL:PROT:SENS:TEMP <Resistance_Value>
23.
Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for
thermal equilibrium, then send this command:
:CAL:PROT:SENS:TEMP 0
24.
Connect the 1000¾ resistor to the INPUT terminals (Figure 2-5).
25.
Send this command to select the 1000¾ range:
:SENS:TEMP:RTD:RANG 1000
26.
Send the following command with the characterized 1000¾ resistance value as the
parameter to calibrate the RTD measurement function:
:CAL:PROT:SENS:TEMP <Resistance_Value>
27.
Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for
thermal equilibrium, then send this command:
:CAL:PROT:SENS:TEMP 0
28.
Connect the DMM and the 1k
Ω
resistor to the INPUT terminals (Figure 2-5).
29.
Send the following command to select a solid-state, current-type temperature sensor:
:SENS:TEMP:TRAN ISS
30.
Send the following command with the characterized 1k
Ω
resistance value as the
parameter:
:CAL:PROT:SENS:TEMP <Resistance_Value>
31.
Note the DMM reading, then send that value as the parameter for the following
command:
:CAL:PROT:SENS:TEMP <DMM_Reading>
32.
Make sure the 1k
Ω
resistor and DMM are still connected to the Model 2510 INPUT
33.
Send the following command to select a solid-state, voltage-type temperature sensor:
:SENS:TEMP:TRAN VSS
2510-902-01 (E - Mar 2010)(Service).book Page 25 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...