2-26
Calibration
Models 2510 and 2510-AT
34.
Note the DMM reading, then send the following command with the DMM reading
value as the parameter to calibrate the current type solid-state sensor:
:CAL:PROT:SENS:TEMP <DMM_Reading>
35.
Short the INPUT terminals with clean copper wire (Figure 2-4). Allow one minute for
thermal equilibrium, then send this command:
:CAL:PROT:SENS:TEMP 0
Step 5: AC ohms calibration
1.
Connect the 100
Ω
resistor to the OUTPUT terminals (see Figure 2-6). Note that the
output will be turned off for the AC ohms calibration step.
2.
Send the following command with the characterized 100
Ω
resistance value as the com-
mand parameter to calibrate the AC ohms function:
:CAL:PROT:RES <Resistance_Value>
Step 6: Voltage source calibration
1.
Disconnect all resistors and test leads from the Model 2510 OUTPUT terminals.
2.
Turn on the output by sending:
:OUTP ON
3.
Send this command to complete voltage source calibration:
:CAL:PROT:SOUR:AUTO
4.
Turn off the output by sending:
:OUTP OFF
Step 7: Program calibration dates
Use following commands to set the calibration date and calibration due date:
:CAL:PROT:DATE
<yyyy>, <mm>, <dd>
(Calibration date)
:CAL:PROT:NDUE
<yyyy>, <mm>, <dd>
(Next calibration due date)
Note that the year, month, and date must be separated by commas.
Step 8: Save calibration constants
Calibration is now complete, so you can store the calibration constants in EEROM by send-
ing the following command:
:CAL:PROT:SAVE
NOTE
Calibration will be temporary unless you send the SAVE command.
Step 9: Lock out calibration
To lock out further calibration, send the following command after completing the calibration
procedure:
:CAL:PROT:LOCK
2510-902-01 (E - Mar 2010)(Service).book Page 26 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...