Models 2510 and 2510-AT
Calibration
2-23
Step 2: Voltage calibration
1.
Connect the DMM to the Model 2510 OUTPUT terminals, as shown in Figure 2-1.
2.
Select the DMM DC voltage function, and enable auto-range.
3.
Send this command to turn on the output:
:OUTP ON
4.
Send the following command to 8.5V:
:SOUR:VOLT 8.5
5.
Note and record the DMM reading, and then send that value as the parameter for the
following commands:
:CAL:PROT:SENS:VOLT <DMM_Reading>
:CAL:PROT:SOUR <DMM_Reading>
For example, if the DMM reading is 8.51V, the correct commands are:
:CAL:PROT:SENS:VOLT 8.51
:CAL:PROT:SOUR 8.51
6.
Send the following command to output 0V:
:SOUR:VOLT 0
7.
Note and record the DMM reading, then send that value as the parameter for the follow-
ing commands:
:CAL:PROT:SENS:VOLT <DMM_Reading>
:CAL:PROT:SOUR <DMM_Reading>
8.
Send the following command to output -8.5V:
:SOUR:VOLT -8.5
9.
Note and record the DMM reading, then send that value as the parameter for the follow-
ing commands:
:CAL:PROT:SENS:VOLT <DMM_Reading>
:CAL:PROT:SOUR <DMM_Reading>
10.
Send this command to turn off the output:
:OUTP OFF
11.
Disconnect the DMM from the OUTPUT terminals.
Step 3: Current and current protection calibration
1.
Connect the 1
Ω
resistor to the Model 2510 OUTPUT terminals, as shown in Figure 2-2.
2.
Send this command to turn on the output:
:OUTP ON
3.
Send the following command with the characterized 1
Ω
resistor value as the command
parameter:
:CAL:PROT:SENS:CURR <Resistance_Value>
For example, if the actual resistor value is 1.01
Ω
, the correct command is:
:CAL:PROT:SENS:CURR 1.01
2510-902-01 (E - Mar 2010)(Service).book Page 23 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...