2-14
Calibration
Models 2510 and 2510-AT
17.
From the THERMISTOR CAL menu, select 10k
Ω
, then press ENTER. The unit will
display the following:
THERMISTOR CAL
Connect 10k
Ω
to temperature
©
ß
sensor and press ENTER.
18.
Connect the 10k
Ω
resistor to the Model 2510 INPUT terminals (Figure 2-3).
19.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 010.0000k
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
20.
Use the EDIT keys to adjust the Model 2510 display to agree exactly with the charac-
terized 10k
Ω
resistance value, then press ENTER. The unit will display the following:
THERMISTOR CAL
Connect a Short to temperature
©
ß
sensor and press ENTER.
21.
Short all four INPUT terminals together using clean copper wire (see Figure 2-4).
Allow one minute for thermal equilibrium.
22.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 00.00000
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
23.
Press ENTER to complete the shorted calibration step, then remove the shorting wires
from the INPUT terminals.
24.
From the THERMISTOR CAL menu, select 100k
Ω
, then press ENTER. The unit will
display the following:
THERMISTOR CAL
Connect 100k
Ω
to temperature
©
ß
sensor and press ENTER.
25.
Connect the 100k
Ω
resistor to the Model 2510 INPUT terminals (Figure 2-3).
26.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 0100.000k
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
27.
Use the EDIT keys to adjust the Model 2510 display to agree exactly with the charac-
terized 100k
Ω
resistance value, then press ENTER. The unit will display the following:
THERMISTOR CAL
Connect a Short to temperature
©
ß
sensor and press ENTER.
28.
Short all four INPUT terminals together using clean copper wire (see Figure 2-4).
Allow one minute for thermal equilibrium.
29.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 00.00000
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
2510-902-01 (E - Mar 2010)(Service).book Page 14 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...