A-2
Specifications
Models 2510 and 2510-AT
HW 3/13/02
Rev. B
2510 TEC SourceMeter
THERMISTOR MEASUREMENT ACCURACY
14
Nominal
Thermistor
Accuracy vs. Temperature
Resistance
0°C
25°C
50°C
100°C
100
Ω
0.021°C
0.035°C
0.070°C
0.27°C
1 k
Ω
0.015°C
0.023°C
0.045°C
0.18°C
10 k
Ω
0.006°C
0.012°C
0.026°C
0.15°C
100 k
Ω
0.009°C
0.014°C
0.026°C
0.13°C
OPEN/SHORTED ELEMENT DETECTION
SOFTWARE LINEARIZATION FOR THERMISTOR AND RTD
COMMON MODE VOLTAGE:
30VDC.
COMMON MODE ISOLATION
: >10
9
Ω
, <1000pF
MAX. VOLTAGE DROP IN INPUT FORCE LEADS:
1 volt.
MAX. SENSE LEAD RESISTANCE:
100
Ω
for rated accuracy.
SENSE INPUT IMPEDANCE:
> 1·10
8
Ω
.
GENERAL
NOISE REJECTION:
SPEED
NPLC
CMRR
12
Normal
1.00
90 dB
SOURCE OUTPUT MODES:
Fixed DC level.
PROGRAMMABILITY:
IEEE-488 (SCPI-1995.0), RS-232, 3 user-definable power-up states plus
factory default and *RST.
POWER SUPPLY:
Nominal 100 to 240VAC rms, 50–60Hz, 90VA.
WARRANTY:
1 year.
EMC:
Conforms to European Union Directive 89/336/EEC, EN 61326-1.
SAFETY:
Conforms to European Union Directive 73/23/EEC, EN 61010-1.
VIBRATION:
MIL-PRF-28800F Class 3 Random Vibration.
WARM-UP:
1 hour to rated accuracies.
DIMENSIONS, WEIGHT:
89mm high
×
213 mm wide
×
370mm deep (3
1
⁄
2
in
×
8
3
⁄
8
in
×
14
9
⁄
16
in).
Bench configuration (with handle & feet):
104mm high
×
238mm wide
×
370mm deep (4
1
⁄
8
in
×
9
3
⁄
8
in
×
14
9
⁄
16
in).
Net Weight:
3.8kg (8.38 lbs).
ENVIRONMENT: Operating:
0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C.
Storage:
–25° to 65°C
The Model 2510 Thermoelectric Cooler Controller is designed to:
• control the power to the TEC to maintain a constant temperature,
current, voltage, or thermistor resistance
• measure the resistance of the TEC
• software PID loop
TEC OUTPUT SPECIFICATIONS
OUTPUT RANGE:
±10 VDC at up to ±5 ADC.
OUTPUT RIPPLE:
<5mV rms.
5
AC RESISTANCE EXCITATION:
±(9.6mA + 190μA).
10, 11
TEC MEASUREMENT SPECIFICATIONS
FUNCTION
1 Year, 23°C ±5°C
Operating
Resistance
1, 6, 7, 8
±(2.0% of rdg + 0.1
Ω
)
Operating Voltage
1, 6
±(0.1% of rdg + 4mV)
Operating Current
6
±(0.4% of rdg + 8mA)
AC Resistance
1, 13
±(0.10% of rdg + 0.02
Ω
)
OPEN SHORTED THERMOELECTRIC DETECTION
LOAD IMPEDANCE:
Stable into 1μF typical.
COMMON MODE VOLTAGE:
30VDC maximum.
COMMON MODE ISOLATION:
>10
9
Ω
, <1500pF.
MAX. SENSE LEAD RESISTANCE:
1
Ω
for rated accuracy.
MAX. FORCE LEAD RESISTANCE:
0.1
Ω
.
THERMAL FEEDBACK ELEMENT SPECIFICATIONS (1 Year, 23°C ±5°C)
Sensor Type
RTD
Thermistor
Solid State
Current
Voltage
100
Ω
1 k
Ω
100
Ω
1 k
Ω
10 k
Ω
100 k
Ω
Output (I
SS
)
Output (V
SS
)
Excitation
9
2.50 mA
833 μA
2.5 mA
833 μA
100 μA
33 μA
2.5 mA
+13.5 V
Compliance
833 μA max
15.75 V max
Nominal Resistance Range
0–250
Ω
0-2.50 k
Ω
0–1 k
Ω
0–10 k
Ω
0–80 k
Ω
0–200 k
Ω
Excitation Accuracy
-2.9%
-2.9%
±2.9%
±2.9%
±2.9%
±2.9%
±12%
±2.9%
Nominal Sensor
Temperature Range
–50° to +250°C
–50° to +250°C –50° to +250°C
–50° to +250°C
–50° to +250°C
–50° to +250°C
–40° to +100°C
–40° to +100°C
Sensor Coefficients
α, β, δ
α, β, δ
A, B, C
A, B, C
A, B, C
A, B, C
Slope & offset
Slope & offset
Measurement Accuracy
±(% rdg + offset)
0.04 + 0.07
Ω
0.04 + 0.4
Ω
0.04 + 0.07
Ω
1
0.04 + 0.4
Ω
1
0.02 + 3
Ω
1
0.04 + 21
Ω
0.03 + 100 nA
0.03 + 500 μV
CONTROL SYSTEM SPECIFICATIONS
SET:
Constant Peltier Temperature
Constant Peltier Voltage
Constant Peltier Current
Constant Thermistor Resistance
CONTROL METHOD:
Programmable software PID loop.
Proportional, Integral, and Derivative gains independently program-
mable.
SETPOINT SHORT TERM STABILITY:
±0.005°C rms.
2, 3
SETPOINT LONG TERM STABILITY:
±0.01°C.
2, 4
SETPOINT RANGE:
–50°C to 225°C.
OVER TEMPERATURE LIMIT:
250°C max.
UNDER TEMPERATURE LIMIT:
–50°C max.
SETPOINT RESOLUTION:
0.001°C, 1mV, 100μA, 0.01% of nominal (25°C)
thermistor resistance.
HARDWARE CURRENT LIMIT:
1.0A to 5.25A ±5%.
SOFTWARE VOLTAGE LIMIT:
±0.5 to 10.5V ±5%.
NOTES
1 With remote voltage sense.
2 With 10k
Ω
thermistor as sensor.
3 Short term stability is defined as
24 hours with Peltier and Model
2510 at 25°C ±0.5°C.
4 Long term stability is defined as
30 days with Peltier and Model
2510 at 25°C ±0.5°C.
5 10Hz to 10MHz measured at 5A
output into a 2
Ω
load.
6 Common mode voltage = 0V
(meter connect enabled, connects
Peltier low output to thermistor
measure circuit ground). ±(0.1%
of rdg + 0.1
Ω
) with meter connect
disabled.
7 Resistance range 0
Ω
to 20
Ω
for
rated accuracy.
8 Current through Peltier > 0.2A.
9 Default values shown, selectable
values of 3μA, 10μA, 33μA, 100μA,
833μA, 2.5mA. Note that tempera-
ture control performance will
degrade at lower currents.
10 AC Ohms is a dual pulsed meas-
urement using current reversals
available over bus only.
11 @23°C ±5°C.
12 For 1k
Ω
unbalance in LO lead.
Minimum amplifier specification.
13 Resistance range 0
Ω
to 100
Ω
for
rated accuracy.
14 Accuracy figures represent the
uncertainty that the Model 2510
may add to the temperature
measurement, not including ther-
mistor uncertainty. These accura-
cy figures are for thermistors with
typical A, B, C constants.
2510-902-01 (E - Mar 2010)(Service).book Page 2 Tuesday, March 16, 2010 2:33 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...