Models 2510 and 2510-AT
Calibration
2-15
30.
Press ENTER to complete the shorted calibration step, then remove the shorting wires
from the INPUT terminals.
31.
Press EXIT to return to the TEMPERATURE CAL menu.
32.
Select RTD, then press ENTER. The unit displays the following:
RTD CAL
Connect 100
Ω
to temperature
©
ß
sensor and press ENTER.
33.
Connect the 100
Ω
resistor to the Model 2510 INPUT terminals (Figure 2-3).
34.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 0.100000k
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
35.
Adjust the display to agree with the value of the 100
Ω
resistor, then press ENTER. The
unit will display the following:
RTD CAL
Connect a Short to temperature
©
ß
sensor and press ENTER.
36.
Short all four INPUT terminals together using clean copper wire (see Figure 2-4).
Allow one minute for thermal equilibrium.
37.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 00.00000
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
38.
Press ENTER to complete the shorted calibration step, then remove the shorting wires
from the INPUT terminals.
39.
Connect the 1k
Ω
resistor to the Model 2510 INPUT terminals (Figure 2-3).
40.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 01.00000k
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
41.
Adjust the display to agree with the value of the 1k
Ω
resistor, then press ENTER. The
unit will display the following:
RTD CAL
Connect a Short to temperature
©
ß
sensor and press ENTER.
42.
Short all four INPUT terminals together using clean copper wire (see Figure 2-4).
Allow one minute for thermal equilibrium.
43.
Press ENTER. The Model 2510 will display the following:
RES VALUE: 00.00000
Ω
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
2510-902-01 (E - Mar 2010)(Service).book Page 15 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...