Calibration
2-5
Calibration menu
Table 2-3 summarizes the main calibration menu selections. To enter the calibration menu,
press the MENU key, select CAL, then press ENTER. Use the EDIT keys to move the cursor
and scroll through menu selections. Press ENTER to select a MENU item.
Unlocking calibration
Before performing calibration, you must first unlock calibration by entering or sending the
calibration password as explained in the following paragraphs.
Unlocking calibration from the front panel
1.
Press the MENU key, choose CAL, and press ENTER. The instrument will display the
following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
©
ß
SAVE LOCK CHANGE-PASSWORD
2.
Select UNLOCK, and then press ENTER. The instrument will display the following:
PASSWORD:
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
3.
Use the
¹
and
ƒ
keys to select the letter or number, and use the
ß
and
©
arrow keys to
choose the position. (Press
ƒ
for letters;
¹
for numbers.) Enter the present password on
the display. (Front panel default: 002510.)
4.
Once the correct password is displayed, press the ENTER key. You can then proceed
with the calibration procedure.
Table 2-3
Calibration menu
Menu selection
Description
UNLOCK
Unlock calibration using password (default: 002510).
EXECUTE
VOLTAGE
CURRENT
I-PROTECTION
TEMPERATURE
AC-OHMS
V-SOURCE
Execute calibration steps.
Voltage calibration.
Current calibration.
Current protection calibration.
Temperature calibration.
AC resistance calibration.
Voltage source calibration.
VIEW-DATES
View calibration dates.
SAVE
Save calibration constants.
LOCK
Lock out calibration.
CHANGE-PASSWORD
Change calibration password.
2510-902-01 (E - Mar 2010)(Service).book Page 5 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...