
4-2
Troubleshooting
Models 2510 and 2510-AT
Introduction
This section of the manual will assist you in troubleshooting and repairing the Model 2510.
Included are self-tests, test procedures, troubleshooting tables, and circuit descriptions. Note
that disassembly instructions are located in Section 5, and component layout drawings are at
the end of Section 6.
Safety considerations
WARNING
The information in this section is intended for qualified service personnel
only. Do not perform these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages that
could cause personal injury or death. Use caution when working with haz-
ardous voltages.
Repair considerations
Before making any repairs to the Model 2510, be sure to read the following considerations.
CAUTION
The PC boards are built using surface mount techniques and require spe-
cialized equipment and skills for repair. If you are not equipped and/or
qualified, it is strongly recommended that you send the unit back to the
factory for repairs or limit repairs to the PC board replacement level.
Without proper equipment and training, you could damage a PC board
beyond repair.
•
Repairs will require various degrees of disassembly; however, it is recommended that
the Front Panel Tests be performed prior to any disassembly. The disassembly instruc-
tions for the Model 2510 are contained in Section 5 of this manual.
•
Do not make repairs to surface mount PC boards unless equipped and qualified to do so
(see previous CAUTION).
•
When working inside the unit and replacing parts, be sure to adhere to the handling pre-
cautions and cleaning procedures explained in Section 5.
•
Many CMOS devices are installed in the Model 2510. These static-sensitive devices
require special handling as explained in Section 5.
•
Whenever a circuit board is removed or a component is replaced, the Model 2510 must
be recalibrated. See Section 2 for details on calibrating the unit.
2510-902-01 (E - Mar 2010)(Service).book Page 2 Tuesday, March 16, 2010 1:04 PM
Содержание TEC SourceMeter 2510
Страница 3: ...2510 902 01 E Mar 2010 Service book Page 2 Tuesday March 16 2010 1 04 PM...
Страница 14: ...1 PerformanceVerification 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 28: ...2 Calibration 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 54: ...3 RoutineMaintenance 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 58: ...4 Troubleshooting 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 76: ...5 Disassembly 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 82: ...6 ReplaceableParts 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 92: ...A Specifications 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 2 33 PM...
Страница 96: ...B CalibrationReference 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 110: ...C CalibrationProgram 2510 902 01 E Mar 2010 Service book Page 1 Tuesday March 16 2010 1 04 PM...
Страница 119: ...2510 902 01 E Mar 2010 Service book Page 4 Tuesday March 16 2010 1 04 PM...