2-40
Model 2790 SourceMeter
®
Switch System Reference Manual
Table 2-5
Command sequence for memory patterns test example
Command sequence
Description
Initial Setup – perform initial setup of instrument, including setting global parameters that apply to all
tests to be executed.
*RST
Reset 2790; opens all channels.
Delay 2
Wait a little for reset to finish.
:FORM:ELEM READ
Send READINGS only when data is requested.
:SYST:AZER:STAT ON or OFF
Enable or Disable A/D Autozero feature. Disabling
increases speed, but A/D accuracy will decrease
over time.
:DISP:ENAB ON or OFF
Turn front panel display on or off. Turning off dis-
play increases speed somewhat.
*** Setup Slot 1 Sources – if the same source settings will be used for all tests, you can set them once as
part of an initial setup. If you must change source levels for different tests, then save the source settings in
the appropriate Memory Pattern for each individual test:
:SOUR:VOLT 500, (@128)
Set V-Source to 500V.
:SOUR:CURR 0.05, (@127)
Set I-Source 50mA.
Note: Allow 250ms maximum settling time for V-source whenever source levels are increased. Fall times
are much slower and settling times up to 1000ms may be required whenever source levels are decreased.
:ROUT:MEM:CLE:ALL
Clear all memory patterns.
Setup Contact Check Test – Memory Pattern Location 1:
:ROUT:MEM:CHAN 1,(@108,115,118)
Specify relays to close.
:ROUT:MEM:DEL 1,0
Set Memory Delay = 0.
:SENS:FUNC 'RES',(@M1)
Select 2-wire ohms measurement function.
:SENS:RES:RANG 100,(@M1)
Select 100 ohm fixed measurement range.
:SENS:RES:NPLC 1,(@M1)
Set A/D integration time to 1 power line cycle,
which is 16.67ms for 60Hz power.
:ROUT:MEM:READ:STAT 1,ON
Enable measurement during scan; during a scan the
measurement is initiated automatically. This does not
apply when a memory pattern is recalled as shown
below. In the latter case, the reading must be initiated
explicitly as also shown later in this example.
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...