Model 2790 SourceMeter
®
Switch System Reference Manual
2-43
Setup Shunt Bar B Test – Memory Pattern Location 6:
:ROUT:MEM:CHAN 6,
(@104,105,117,118,121)
Specify relays to close.
:ROUT:MEM:SOUR:LEV 6,0.05,(@127)
Set Isrc to 50mA for M6.
:ROUT:MEM:DEL 6,0
Set Memory Delay = 0.
:SENS:VOLT:NPLC 1,(@M6)
Set A/D integration time to 1 PLC.
:CALC1:FORM S1I,(@M6)
Select S1IOHMS math function; selecting this
function automatically selects DCV measurement
function and the 1V measurement range.
:CALC1:STAT ON,(@M6)
Enable the math function.
:ROUT:MEM:READ:STAT 6,ON
Enable measurement during scan; during a scan the
measurement is initiated automatically. This does
not apply when a memory pattern is recalled as
shown below. In the latter case, the reading must be
initiated explicitly as also shown below.
Setup Open Circuit Test Voltage Verification – Memory Pattern Location 7:
:ROUT:MEM:CHAN 7,
(@109,112,122,113,118,123,121)
Specify relays to close (109 and 112 are shunt bars).
:ROUT:MEM:SOUR:LEV 7,500,(@128)
Set Vsrc to 500V for M7. Since source level is the
same as in Initial Setup routine, this command can
be left out.
:ROUT:MEM:DEL 7,0.05
Set Total Memory Delay = 50ms, allow 50ms settling
time after changing load on V-source; if also change
source level, include an additional 250ms for total of
300ms.
:SENS:FUNC 'VOLT',(@M7)
Select DC volts measurement function.
:SENS:VOLT:RANG 1000,(@M7)
Select 1000V fixed measurement range.
:SENS:VOLT:NPLC 1,(@M7)
Set A/D integration time to 1 PLC.
:ROUT:MEM:READ:STAT 7,ON
Enable measurement during scan; during a scan the
measurement is initiated automatically. This does
not apply when a memory pattern is recalled as
shown below. In the latter case, the reading must be
initiated explicitly as also shown below.
Table 2-5 (cont.)
Command sequence for memory patterns test example
Command sequence
Description
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...