5-4
Model 2790 SourceMeter
®
Switch System Reference Manual
Remote programming — rel
Rel commands
The rel commands to set range are listed in
. Additional information on these
commands follow the table.
klqb
Query commands are not included in
. All commands for the SENSe
subsystem are provided in
Table 5-1
Rel commands
Commands
1
Description
Default
Rel commands for DCV
[SENSe[1]]
Optional root command.
:VOLTage[:DC]:REFerence <n> [, <clist>]
Specify rel value; <n> = -1010 to 1010
(V).
0
:VOLTage[:DC]:REFerence:STATe <b>
[, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:VOLTage[:DC]:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for ACV
[SENSe[1]]
Optional root command.
:VOLTage:AC:REFerence <n> [, <clist>]
Specify rel value; <n> = -757.5 to 757.5
(V).
0
:VOLTage:AC:REFerence:STATe <b> [, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:VOLTage:AC:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for DCI
[SENSe[1]]
Optional root command.
:CURRent[:DC]:REFerence <n> [, <clist>]
Specify rel value; <n> = -3.1 to 3.1 (A).
0
:CURRent[:DC]:REFerence:STATe <b>
[, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:CURRent[:DC]:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Rel commands for ACI
[SENSe[1]]
Optional root command.
:CURRent:AC:REFerence <n> [, <clist>]
Specify rel value; <n> = -3.1 to 3.1 (A).
0
:CURRent:AC:REFerence:STATe <b>
[, <clist>]
Enable/disable rel; <b> = ON or OFF.
OFF
:CURRent:AC:REFerence:ACQuire [, <clist>]
Use input signal as rel value.
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...