4-6
Model 2790 SourceMeter
®
Switch System Reference Manual
Digits
The DIGITS key sets display resolution for the Model 2790 from 3 to 6 digits. From the
front panel, setting digits for one function affects all the other functions. For example if
you set DCV for 3 digits, the other functions will also set to 3 digits. For remote pro-
gramming, each mainframe input function can have its own unique digits setting.
Digits has no effect on the remote reading format. The number of displayed digits does not
affect accuracy or speed. Speed is set by the RATE key.
Setting display resolution
— To set display resolution, press the DIGITS key until the
desired number of digits is displayed.
Scanning (7702 only)
When a simple scan using a 7702 module is configured, the present digits setting will
apply to all channels in the scan. When an advanced scan is configured, each channel can
have its own unique digits setting. Details to configure and run a scan are provided in
.
For remote programming, the <clist> parameter is used to configure channels for a scan.
Remote programming — digits
Digits commands
The commands to control display resolution (digits) are listed in
. Additional
information on these commands follow the table.
klqb
Query commands are not included in
. All commands for the SENSe
subsystem are provided in
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...