1-36
Model 2790 SourceMeter
®
Switch System Reference Manual
Exercise 5 — Trigger and return a single reading
Exercise 6 — Trigger and return multiple readings
Trigger controlled measurements
— The instrument is typically used in a non-
continuous trigger mode. In this mode, commands are used to trigger one or more read-
ings. After the specified number of readings are completed, the measurement process
stops.
Exercise 5 in
provides a command sequence to trigger and return one reading.
Exercise 6 in
provides a command sequence to trigger and return multiple
readings.
Figure 1-7
Exercise 5 — Trigger and return a single reading
INIT:CONT OFF
TRIG:COUN 1
SAMP:COUN 1
Place 2790 in
non-continuous
trigger state
Set 2790 to perform
one measurement
1. If a MATH function is enabled, the result of the calculation will be returned
(MATH functions are covered in Section 5).
3. FETCh?, CALC:DATA? and DATA? do not trigger readings. They simply return the last
reading. If you again send one of these commands before triggering a new reading,
the old reading will be returned.
Return Basic Reading
3,4
INIT
Trigger Reading
READ?
2. If there is no MATH function enabled, FETCh? and CALC:DATA? will return the basic
reading.
Trigger and Return
Reading
1
Trigger Configuration
Trigger and Return
Reading
DATA?
OR
DATA:FRESh?
Return result of MATH
calculation
1, 2, 3
4. DATA:FRESh? can only be used once to return the same reading. Sending it again
without first triggering a new reading will cause error -230 (data corrupt or stale).
FETCh?
OR
CALC:DATA?
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...