Model 2790 SourceMeter
®
Switch System Reference Manual
5-21
Remote programming — ratio and channel average
Ratio and channel average commands
The ratio and channel average commands are listed in
. Details on these com-
mands follow the table.
klqb
Queries are not included in
. All the ratio and channel average com-
.
Enabling/disabling ratio or channel average
As with front panel operation, enabling ratio disables channel average and conversely,
enabling channel average disables ratio.
Ratio and channel average delay
:RATio:DELay or :CAVerage:DELay sets the delay between the two channel measure-
ments for the enabled calculation. This delay is applied after the trigger delay in the trigger
model (see
for details). This delay cannot be set from the front panel. The 0.5s
default delay keeps the relays from cycling too fast. Setting a shorter delay may shorten
the life of the relays.
It does not matter which of the two commands you use to set the delay. The set delay
affects both ratio and channel average.
Table 5-3
Ratio and channel average commands
Commands*
Description
Def
[SENSe[1]]
Optional root command.
:RATio[:STATe] <b> [, <clist>]
Enable/disable ratio; <b> = ON or OFF.
OFF
:RATio:DELay <NRf> [, <clist>]
Set delay (in secs); <NRf> = 0 to 99999.999.
0.5
:CAVerage[:STATe] <b> [, <clist>]
Enable/disable channel average; <b> = ON or OFF.
OFF
:CAVerage:DELay <NRf> [, <clist>]
Set delay (in secs); <NRf> = 0 to 99999.999.
0.5
Channel list parameter (7702):
<clist> = (@SCH)
where: S
= Mainframe slot number (1 or 2)
CH = Switching module channel number (must be 2 digits)
Examples:
(@101) = Slot 1, Channel 1
(@101, 203) = Slot 1, Channel 1 and Slot 2, Channel 3
(@101:110) = Slot 1, Channels 1 through 10
*The <clist> parameter is used to configure one or more 7702 channels for a scan.
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...