Model 2790 SourceMeter Switch System Reference Manual
9-11
Setting digital output
The OUTPUT menu (shown in
) is used to control and configure digital outputs.
Menu items for the digital output include:
•
DOUTPUT
— Use to enable (ON) or disable (OFF) the digital outputs.
•
PULSE
— Use to enable (YES) or disable (NO) the pulse option for the digital
outputs.
klqb
The factory default pulse time is 2ms maximum. Using remote programming,
pulse time can be set from 0.001 to 99999.999 sec. It cannot be set from the front
panel.
•
LSENSE
— Use to select the logic sense: active HIGH or active LOW. With active
high selected, an output will be at approxi5V when a reading is at or
exceeds the limit. Conversely, with active low selected, an output will be at 0V
when a reading reaches or exceeds the limit.
•
MASTR LATCH
— Use to enable (Y) or disable (N) the master limit latch. When
enabled, the master limit remains latched when a reading limit is reached or
exceeded. When disabled, the master limit line releases immediately when the
reading is inside all four limits.
Perform the following steps to enable and configure digital outputs:
1.
Press SHIFT and then OUTPUT.
2.
If the digital output is already on (DOUTPUT: ON), proceed to step 3. Otherwise,
press
to move the cursor to the right, press the
or
key to display “ON,”
and press ENTER.
3.
Use the
key to display the master limit latch (MASTR LATCH) setting: N (no)
or Y (yes).
4.
If you want to retain the present master limit setting, proceed to step 5. Otherwise,
press
to move the cursor to the right, press
or
key to display “Y” or “N,”
and press ENTER.
Table 9-1
OUTPUT menu
Menu item
Setting
Description
DOUTPUT
ON or OFF
Enable/disable digital outputs.
PULSE
YES or NO
Enable/disable digital pulse
output.
LSENSE
HIGH or LOW
Select logic sense.
BEEP
NEVER, INSIDE, or OUTSIDE Set beeper for limits
” for details).
MASTR LATCH
Y or N
Enable/disable master limit latch.
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...