17-26
Model 2790 SourceMeter
®
Switch System Reference Manual
9.
Set the generator to output a 1V RMS, 1kHz sine wave, then send the following
command:
:CAL:PROT:AC:STEP15 1E3
10.
Send the following commands to set calibration dates, save calibration, and lock
out calibration:
:CAL:PROT:DATE <yyyy>, <mm>, <dd>
:CAL:PROT:NDUE <yyyy>, <mm>, <dd>
:CAL:PROT:SAVE
:CAL:PROT:LOCK
Model 7751, 7752, and 7753 calibration
The following procedures calibrate the current source of the Model 7751/7752/7753 module.
The voltage source and I/V converter do not need calibration.
t^okfkd
For the Model 7751, 7752, and 7753 modules, the maximum
common-mode voltage (voltage between any module terminal
and chassis ground) is 42V. Exceeding this value may cause a
breakdown in insulation, creating a shock hazard.
klqb
The source calibration procedure below needs to be performed only once every
two years for the module sources to meet their source output accuracy specifica-
tions. However, you must calibrate the Model 2790 mainframe at least once a
year to meet Model 7751/7752/7753 resistance accuracy specifications.
Make sure the interlock is enabled.
Recommended test equipment
In order to calibrate the Model 7751/7752/7753, you will need the equipment summarized
in
.
Table 17-15
Recommended equipment for
Model 7751/7752/7753 calibration
HP3458A Digital Multimeter:
50mA: ±35ppm
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...