Model 2790 SourceMeter
®
Switch System Reference Manual
6-9
Table 6-1
Buffer commands
Command
Description
Default
1
Ref
SYSTem:TIME <hr, min, sec>
Set clock time in 24-hour format.
a
SYSTem:DATE <yr, mo, day>
Set clock date; yr specified as 20xx.
b
SYSTem:TSTamp:TYPE
<name>
Select timestamp; <name> = RELative or RTCLock.
REL
c
SYSTem:TSTamp:TYPE?
Query timestamp type that will be used for
the next buffer storage.
c
TRACe:TSTamp:TYPE?
Query timestamp type for readings presently in buffer.
c
TRACe:CLEar
Clear the buffer immediately.
d
TRACe:CLEar:AUTO <b>
Enable/disable buffer auto-clear; <b> = ON or OFF.
ON
d
TRACe:FREE?
Query bytes available and bytes in use.
e
TRACe:POINts <NRf>
Specify size of buffer; <NRf> = 2 to 55000.
100
f
TRACe:TSTamp:FORMat
<name>
Select timestamp format; <name> = ABSolute or
DELTa.
ABS
g
TRACe:FEED <name>
Select source of readings; <name> = SENSe[1],
CALCulate[1] or NONE.
CALC
h
TRACe:FEED:CONTrol
<name>
Set buffer control; <name> = NEVer, NEXT, or
ALWays.
NEV
h
TRACe:DATA?
Read all readings in the buffer.
i
TRACe:DATA:SELected?
<start>, <count>
Specify readings to be returned; <start> = starting
point, <count> = number of readings.
j
TRACe:NEXT?
Query buffer location for next stored reading.
j
TRACe:NOTify <NRf>
Specify number of stored readings that will set Trace
Notify bit (B6) of measurement event register;
<NRf> = 2 to 109999 (must be less than
TRACe:POINts value).
k
FORMat:ELEMents <item list> Specify elements for TRACe:DATA? response
message; <item list> = READing, CHANnel,
UNITs, RNUMber, and TSTamp.
(Note 2)
l
CALCulate2:FORMat <name> Select buffer statistic; <name> = MINimum,
MAXimum, MEAN, SDEViation, PKPK, or
NONE.
NONE
m
CALCulate2:STATe <b>
Enable/disable statistic calculation; <b> = ON or
OFF.
OFF
m
CALCulate2:IMMediate
Calculate data in buffer.
m
CALCulate2:IMMediate?
Calculate data and read result.
m
CALCulate2:DATA?
Read the selected buffer statistic.
m
Notes:
1. SYSTem:PRESet and *RST have no effect on TRACe commands. The listed defaults for TRACe commands are set at the factory.
2. The SYSTem:PRESet and *RST default is READ, UNIT, RNUM, and TST.
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...