Model 2790 SourceMeter
®
Switch System Reference Manual
1-33
Figure 1-6
Simple scan operation
For remote programming, the following commands are used for simple scanning:
ROUTe:SCAN <clist>
' Define scan list*.
TRIGger:COUNt <n>
' Specify number of scans (1 to 55000 or
INFinity).
SAMPle:COUNt <n>
' Specify number of channels to scan (1 to
55000).
ROUTe:SCAN:LSELect <name>
' Enable (INT) or disable (NONE) scan.
* Any valid switching module channel can be included in the scan list. Make sure to list
them from the lowest numbered channel to the highest. For example, to scan channels 1
through 8 of a Model 7702 installed in slot 2, send the following command to define the
scan list:
ROUTe:SCAN (@201:208)
MIN CHAN: XXX
INT: SIMPLE
Display SIMPLE option
and press ENTER
Specify minimum
channel (XXX) and
press ENTER
SHIFT
SCAN
STEP
CONFIG
Press SHIFT
Press CONFIG (STEP)
MAX CHAN: YYY
Specify maximum
channel (YYY)
and press ENTER
TIMER? NO/YES
Display NO or YES
and press ENTER
YES
NO
xxH:xxM:xx.xxxS
Set timer interval in
hr:min:sec format and
press ENTER
RDG CT:xxxxxx
Specify reading count
and press ENTER
SCAN
STEP
Press STEP or SCAN to start
scan
Timer interval specifies time
between scans.
Reading count:
Specifies number of scans
to be performed.
Specifies number of readings
to store in buffer.
Timer interval specifies time
between scanned channels.
Reading count:
Specifies number of channels to be
scanned.
Specifies number of readings to store
in buffer.
Step 1. Configure simple scan:
Step 2. Run simple scan:
SHIFT
SCAN
STEP
HALT
Press SHIFT
Press HALT (SCAN)
Step 3. Disable scan mode:
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...