13-8
Model 2790 SourceMeter
®
Switch System Reference Manual
MEASure:<function>? [<rang>], [<res>], [<clist>]
MEASure:VOLTage[:DC]? [<rang>], [<res>], [<clist>]
Measure DCV
MEASure:VOLTage:AC? [<rang>], [<res>], [<clist>]
Measure ACV
MEASure:CURRent[:DC]? [<rang>], [<res>], [<clist>]
Measure DCI
MEASure:CURRent:AC? [<rang>], [<res>], [<clist>]
Measure ACI
MEASure:RESistance? [<rang>], [<res>], [<clist>]
Measure
Ω
2
MEASure:FRESistance? [<rang>], [<res>], [<clist>]
Measure
Ω
4
MEASure:FREQuency? [<rang>], [<res>], [<clist>]
Measure FREQ
MEASure:PERiod? [<rang>], [<res>], [<clist>]
Measure PERIOD
MEASure:TEMPerature? [<clist>]
Measure TEMP
MEASure:CONTinuity? [<clist>]
Measure CONT
Parameters
<rang>
= Range parameter for the specified function. For example for
DCV, range parameter value 10 selects the 10V range. See the
“NOTES” that follow
for additional information.
<res>
= 0.1
i.e., 100.0 V (3 digits)
0.01
i.e., 10.00 V (3 digits)
0.001
i.e., 1.000 V (3 digits)
0.0001
i.e., 1.0000 V (4 digits)
0.00001
i.e., 1.00000 V (5 digits)
0.000001
i.e., 1.000000 V (6 digits)
The resolution of the <res> parameter value and the selected
range sets the number of display digits. As shown above, with
the 100V range selected and <res> = 0.1, a 100V reading will
be displayed as 100.0 V (3 digits).
The display will default to 3 digits when using parameter
values that attempt to set the display below 3½ digits. For
example, a 10V reading using <res> = 0.1 for the 10V range is
displayed as 10.00 V, not 10.0 V.
A command using parameter values that attempt to set the
display above 7 digits is ignored and generates error -221
(settings conflict error).
for additional
information.
<clist>
= Single channel only. When included, this is the channel to be
closed and measured.
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...