3-10
Model 2790 SourceMeter
®
Swich System Reference Manual
Model 7702 switching module
Connections for the 7702 switching module are shown in
. For basic DCV and
ACV measurements (
A and B), channels 1 through 40 can be used.
Ratio and channel average calculations
— Ratio calculates the reading ratio of two
channels, while channel average calculates the reading average of two channels. For these
calculations, paired switching channels are used. Primary channels 1 through 20 are paired
to channels 21 through 40 (channel 1 paired to channel 21, channel 2 paired to channel 22,
and so on). As shown in
C, one DC voltage source is connected to a primary
channel (i.e., 104) and the other source is connected to its paired channel (i.e., 124).
klqb
The ratio and channel average calculations are covered in
Figure 3-3
DCV and ACV connections using Model 7702 switching module
AC Voltage
Source
H
L
CH 1-40
Model 7702
Switching
Module
DC Voltage
Source
H
L
CH 21-40
DC
Voltage
Source
A) DCV connections
B) ACV connections
C) Ratio and channel average connections (DCV)
CAUTION Maximum input: 300VDC or RMS,
1A switched, 60W, 125VA maximum.
NOTE
The low connections for channels 1 through 20 do not
need to be referenced to the low connections for
channels 21 through 40.
H
L
CH 1-40
Model 7702
Switching
Module
H
L
CH 1-20
Model 7702
Switching
Module
DC
Voltage
Source
Содержание 2790
Страница 3: ......
Страница 7: ......
Страница 105: ...2 48 Model 2790 SourceMeter Switch System Reference Manual...
Страница 245: ...7 40 Model 2790 SourceMeter Switch System Reference Manual...
Страница 265: ...8 20 Model 2790 SourceMeter Switch System Reference Manual...
Страница 305: ...10 22 Model 2790 SourceMeter Switch System Reference Manual...
Страница 328: ...12 CommonCommands...
Страница 336: ...13 SCPISignalOriented MeasurementCommands...
Страница 345: ...13 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 355: ...14 10 Model 2790 SourceMeter Switch System Reference Manual...
Страница 356: ...15 SCPIReferenceTables...
Страница 419: ...16 38 Model 2790 SourceMeter Switch System Reference Manual...
Страница 461: ...A 2 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 462: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 3...
Страница 463: ...A 4 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 464: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 5...
Страница 465: ...A 6 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 466: ...Model 2790 SourceMeter Switch System Reference Manual Specifications A 7...
Страница 471: ...A 12 Specifications Model 2790 SourceMeter Switch System Reference Manual...
Страница 472: ...B StatusandErrorMessages...
Страница 479: ...B 8 Model 2790 SourceMeter Switch System Reference Manual...
Страница 493: ...C 14 Model 2790 SourceMeter Switch System Reference Manaul...
Страница 509: ...E 6 Model 2790 SourceMeter Switch System Reference Manual...
Страница 510: ...F IEEE 488BusOverview...
Страница 525: ...F 16 Model 2790 SourceMeter Switch System Reference Manual...
Страница 526: ...G CalibrationReference...